24
MODEL SOLU COMP II
SECTION 5.0
PROGRAMMING THE ANALYZER
5.3.3. Procedure: Configure Outputs.
To choose a menu item, move the cursor to the item and press ENTER.
To store a number or setting, press ENTER.
1.
Press MENU. The main menu screen appears. Choose
Program
.
2.
Choose
Outputs
.
3.
Choose
Output Configure
.
4.
Choose
Output1
or
Output2
.
5.
Choose
Measurement
or
Temp
.
Measurement
means ozone.
6.
Make the appropriate settings:
a. Choose
4-20 mA
or
0-20 mA
.
b.
Choose
Yes
or
No
for output dampening.
c.
Choose
Linear
or
Log
output.
7.
The display returns to the
Output Config?
screen. Select the other out-
put or press EXIT to return to the previous screen. To return to the main
display, press MENU followed by EXIT.
5.3.4. Procedure: Assigning Values to the Low and High Current Outputs (Output Ranging)
To choose a menu item, move the cursor to the item and press ENTER.
To store a number or setting, press ENTER.
1.
Press MENU. The main menu screen appears. Choose
Program
.
2.
Choose
Outputs
.
3.
Choose
Output Range
. Choose
Output1
or
Output2
.
4.
Make the appropriate settings.
a.
Assign a value to the low current (
0 mA
or
4 mA
) output.
b.
Assign a value to the high current (
20 mA
) output.
5.
The display returns to the
Output Range
screen. Select the other output
or press EXIT to return to the previous screen. To return to the main dis-
play, press MENU followed by EXIT.
Calibrate
Hold
P
Pr
ro
og
gr
ra
am
m
Display
Output Config?
O
Ou
ut
tp
pu
ut
t1
1
Output2
OutM is for?
M
Me
ea
as
su
ur
re
em
me
en
nt
t
Temp
Output Range
O
Ou
ut
tp
pu
ut
t
C
Co
on
nf
fi
ig
gu
ur
re
e
O
Ou
ut
tp
pu
ut
ts
s
Alarms
Measurement
>>
Calibrate
Hold
P
Pr
ro
og
gr
ra
am
m
Display
O
Ou
ut
tp
pu
ut
t
R
Ra
an
ng
ge
e
Output Configure
O
Ou
ut
tp
pu
ut
ts
s
Alarms
Measurement
>>