2.3
Simulator status
The status is defined by the halogen light.
Normal status
When switching to normal status, the equipment will operate for a maximum period of 60
seconds, following which the equipment cannot be operated for a minimum period of 30
seconds.
Fault status
When the battery is lower than 12.3 Vdc the lamp won't be activated.
2.4
Charge the battery
This device uses lithium ion batteries as a rechargeable power source. When the batteries
are fully charged, under normal conditions (room temperature), the device operates at
least 1,000 times without restarting. When the voltage from the batteries is lower than
the required operational level, the device will not operate.
To charge the battery:
Procedure
1. Place the device on a table in a safe area not exceeding 104 °F (40 °C).
2. Release the locking screw.
See
3. Unscrew the battery compartment cover counter-clockwise.
See
.
4. Unscrew the locking disc clockwise using a 1.5 mm hex key.
See
.
5. Pull out the battery from the flame simulator.
WARNING
The battery must be charged only after being removed from the flame simulator.
6. Connect the battery to the charger. Verify that the charger is the one supplied with
the flame simulator.
7. Let the battery charge for a maximum of two to three hours.
8. Disconnect the charger.
9. Insert the battery into the flame simulator.
The LED will flash twice to indicate that the battery is fully charged.
10. Screw in the locking disc using a 1.5 mm hex key.
See
.
11. Screw in the battery compartment cover.
Operation
Reference Manual
May 2022
00809-0800-4975
12
Emerson.com/Rosemount
Summary of Contents for Rosemount FS-UVIR-975
Page 1: ...Reference Manual 00809 0800 4975 Rev DD May 2022 Rosemount FS UVIR 975 ...
Page 4: ...4 ...
Page 6: ...Contents Reference Manual May 2022 00809 0800 4975 6 Emerson com Rosemount ...
Page 14: ...Operation Reference Manual May 2022 00809 0800 4975 14 Emerson com Rosemount ...
Page 18: ...Maintenance Reference Manual May 2022 00809 0800 4975 18 Emerson com Rosemount ...
Page 20: ...Ordering information Reference Manual May 2022 00809 0800 4975 20 Emerson com Rosemount ...
Page 26: ...Product certifications Reference Manual May 2022 00809 0800 4975 26 Emerson com Rosemount ...