Figure 6-7: Select gas Screen
7. Use
and
to move the cursor until the gas that you wish to span calibrate is
highlighted by the cursor. Press
.
The
Select type
screen (
Figure 6-8: Select Type Screen
8. Use
and
to move the cursor until
SPAN
). To calibrate,
press
.
The
Mode selection
screen (
Gas calibration
Reference Manual
September 2019
00809-0100-5800
86
Emerson.com/Rosemount