Figure 6-9: Mode selection Screen
9. Press
to perform a manual calibration.
The
Span input
screen (
) opens.
Figure 6-10: Span input Screen
10. Use
and
to highlight each digit in turn; then use
and
to increase or
decrease the value until the concentration displayed matches the cylinder you are
using. Press
to proceed to the next step.
NOTICE
Enter the cylinder concentration in ppm.
The
Begin Calibrate span
screen (
Reference Manual
Gas calibration
00809-0100-5800
September 2019
Rosemount CT5800
87