Calibration Curve
Figure A-5:
This is the calibration curve for the instrument, opposite in shape to the transmissivity
curve shown in
. Characteristics of the 4.7 µm filter fix each curve, and the curves
cannot change. Rosemount analyzers make full use of this scale shape to provide an easily
attainable calibration point.
It is not necessary to calculate K because we know that when the constant K is correct (Y =
0 when the CO level = 0) any drift in the measurement can only be due to a change in some
optical or electronic gain. You can always correctthe drift by setting Y to zero when the CO
level is zero.
In practice, however, it is not always possible to produce a zero CO level. But if we consider
the calibration curve, we can see that:
if Y = 0 when CO = 0
then Y = a when CO = b
We can also see from
that at high CO levels the parameter Y becomes
completely insensitive to changing CO levels in the duct, such that:
Y = a when CO > b
By making Y = a when CO > b and Y = 0 when CO - 0, you can eliminate all errors.
You can introduce a gas cell containing pure carbon monoxide can be introduced into the
IR beam at the source. This cell represents a value of 10,000 ppm-meters and provides a
reference point for the calibration of the instrument. Any further CO in the duct will have
negligible effect on the reference point because the calibration curve is flat at these high
concentrations of CO. Well designed cross-duct analyzers introduce this gas cell every few
seconds to continuously check and (if necessary) modify the zero position.
Theory of operation
112
Rosemount CCO 5500
Summary of Contents for Rosemount CCO 5500
Page 4: ......
Page 8: ...Contents iv Rosemount CCO 5500 ...
Page 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Page 32: ...Install 24 Rosemount CCO 5500 ...
Page 88: ...Operation 80 Rosemount CCO 5500 ...
Page 98: ...Maintenance 90 Rosemount CCO 5500 ...
Page 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Page 110: ...Spare parts 102 Rosemount CCO 5500 ...
Page 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Page 123: ...Return equipment to the factory Reference Manual 115 ...