Safety Instructions
HOW TO STAY IN COMPLIANCE WITH THE
EUROPEAN DIRECTIVE 94/9/EC ("ATEX") WHEN PERFORMING GAS
ANALYSIS WITHIN A FLAMEPROOF ENCLOSURE.
Special conditions apply to using a
fl
ameproof enclosure analyzer under the scope of the
"European Directive for Equipment used in Explosive Atmosphere" (Directive 94/9/EC;
ATEX). To stay compliant to the directive consider the following clari
fi
cation sheet released
by the European ATEX Noti
fi
ed Body Group (see next page):
HEAVY INSTRUMENT
The analyzer model X-STREAM X2FD, to which this manual relates, intended
to be wall mounted and/or outdoor installed, weighs up to approx. 63 kg
(139 lbs), depending on included options!
Use two people and/or suitable tools for transportation and lifting these
instruments!
Take care to use anchors and bolts speci
fi
ed to be used for the weight of
the units!
Take care the wall or stand the unit is intended to be installed at is solid
and stable to hold the units!
HIGH TEMPERATURES
While working at internal components hot surfaces may be accessible, even
after the instrument has been disconnected from power!
EXPLOSION HAZARD BY HOT COMPONENTS
Temperatures inside an analyzer for VAPOR RECOVERY applications
exceed the analyzer‘s temperature classi
fi
cation for hazardous areas!
Special conditions apply to handling and operating this analyzer, consider
the safety instructions at the beginning of this manual!