SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
Explosion-Proof:
Class I, Div. 1, Groups B-D
Class II, Div. 1, Groups E-G
Class III, Div. 1
Class I, Groups B-D
Class II, Groups E-G
Class III
Tamb = 65°C max
1.3 SPECIFICATIONS
—
OxyGEN
Measurement Range:
0-99 ppm (mg/L), 0-200%
saturation
Resolution:
0.01 ppm, 0.1 ppb for 499A TrDO sen-
sor
Temperature correction for membrane perme-
ability:
automatic between 0 and 50°C (can be
disabled)
Calibration:
air calibration (user must enter baro-
metric pressure) or calibration against a stan-
dard instrument
RECOMMENDED SENSORS — OxyGEN:
Model 499A DO-54 for ppm level
Model 499A TrDO-54 for ppb level
Hx438 and Gx448 steam-sterilizable oxygen sensors
1.4 SPECIFICATIONS
—
FREE CHLORINE
Measurement Range:
0-20 ppm (mg/L) as Cl
2
Resolution:
0.001 ppm (Autoranges at 0.999 to
1.00 and 9.99 to 10.0)
Temperature correction for membrane perme-
ability:
automatic between 0 and 50°C (can be
disabled)
pH Correction:
Automatic between pH 6.0 and
9.5. Manual pH correction is also available.
Calibration:
against grab sample analyzed using
portable test kit.
RECOMMENDED SENSOR — FREE CHLORINE:
Model 499A CL-01-54
SPECIFICATIONS — pH
Application:
pH measurement available with free
chlorine only
Measurement Range:
0-14 pH
Resolution:
0.01 pH
Sensor Diagnostics:
Glass impedance (for broken
or aging electrode) and reference offset.
Reference impedance (for fouled reference
junction) is not available.
Repeatability:
±0.01 pH at 25°C
RECOMMENDED SENSOR — pH
:
Use Model 399-09-62, 399-14, or 399VP-09.
See pH sensor product data sheet for complete order-
ing information.
1.5 SPECIFICATIONS
—
TOTAL CHLORINE
Measurement Range:
0-20 ppm (mg/L) as Cl
2
Resolution:
0.001 ppm (Autoranges at 0.999 to
1.00 and 9.99 to 10.0)
Temperature correction for membrane perme-
ability:
automatic between 5 and 35°C (can be
disabled)
Calibration:
against grab sample analyzed using
portable test kit.
RECOMMENDED SENSOR — TOTAL CHLORINE:
Model 499A CL-02-54 (must be used with SCS 921)
1.6 SPECIFICATIONS
—
OzONE
Measurement Range:
0-10 ppm (mg/L)
Resolution:
0.001 ppm (Autoranges at 0.999 to
1.00 and 9.99 to 10.0)
Temperature correction for membrane perme-
ability:
automatic between 5 and 35°C (can be
disabled)
Calibration:
against grab sample analyzed using
portable test kit.
RECOMMENDED SENSOR — OzONE:
Model 499A OZ-54
1.7
SPECIFICATIONS — PERCENT
OxyGEN IN GAS
Measurement Range:
0-25% Oxygen
Resolution:
0.1% - TBD
Calibration:
air calibration (automatic meas
urement of barometric pressure with internal
pressure sensor) - TBD
Sample Pressure:
0 to 50 PSIG
Sample Temperature:
32 to 110°F
RECOMMENDED SENSOR — PERCENT
OxyGEN IN GAS
Model 4000 Percent Oxygen Sensor
Summary of Contents for Rosemount Analytical HART 5081-A-HT
Page 22: ...FIGURE 4 2 FM Intrinsically Safe Installation Label 16 ...
Page 25: ...FIGURE 4 4 CSA Intrinsically Safe Installation Label 19 ...
Page 26: ...FIGURE 4 5 CSA Intrinsically Safe Installation 1 of 2 20 ...
Page 27: ...FIGURE 4 5 CSA Intrinsically Safe Installation 2 of 2 21 ...
Page 108: ...FIGURE 4 Wiring to Model 5081 102 MODEL 4000 WIRING ...
Page 110: ...104 ...