1. Press PROG on the remote controller.
2. Press NEXT until the
dEFAULt
appears in the display. Press ENTER.
3. Use
é
or
ê
to toggle between
nO
and
yES
. With
yES
showing, press ENTER to return
to factory default settings.
PROGRAM
tYPE
EXIT
ENTER
02
PROGRAM
dISPLAY
EXIT
NEXT ENTER
MODEL 5081-A
SECTION 7.0
PROGRAMMING
7.6 FACTORy DEFAULT
7.6.1 Purpose
This section describes how to erase ALL user-defined configuration settings and return the transmitter to factory default
settings.
7.6.2 Procedure
7.7 HART
For more information, see Section 6.3.2.
41