6. Follow the transmitter prompts to complete the calibration.
Note
Refer to the manual for the transmitter you are using (Rosemount
56, 1056, 5081, or 1066).
7. After calibration, go to the
Diagnostics
menu and check the
sensitivity.
The sensitivity should be between 300 and 700 nA/ppm. For more
information, refer to the transmitter manual.