4
Specifications
Measurements and
ranges
pH: 0 - 14
Oxidation reduction potential (ORP): -1500
mV to +1500 mV
Percent linearity over pH
range
Range
Linearity
0-7
97%
1-7
98%
4-7
98%
7-10
99%
7-12
97%
7-13
96%
7-14
95%
Materials of construction
Sensor body
Ryton
®
polyphenylene sulfide (PPS)
O-ring
EPDM
ORP electrode
Glass, platinum
Solution ground
Stainless steel
Reference junction
PFTE
Maximum pressure
100 psig (790 kPa) at 212 °F (100 °C)
CRN rating: 60 psig (413.7 kPa) to 212 °F (100 °C)
Operating temperature
14 to 212 °F (-10 to 100 °C)
Automatic temperature compensation: 14 to 212 °F (-10 to 100 °C)
Conductivity
Responds to changes in pH at a minimum conductivity of 0.1 µS/cm when
used with the low flow panel (PN SQP10077-LQD). The sample flow rate
must be controlled to 2 gph (7.6 L/hr).