2.3
Electrical Installation
Make electrical connections as shown on Figures 2-3 using the following guidelines:
1.
Pay particular attention to the analyzer or transmitter model number when following
details on the wiring diagrams to ensure that the connections are made to the proper
terminals.
2.
Use a high quality four conductor shielded instrument cable such as Belden 8722. This is
available from Rosemount as Part Number 9200000.
3.
Maximum sensor to analyzer distance is: (Integral preamplifier in sensor) 1054, 1054A
pH/ORP, 1054B pH/ORP and 2054 pH - 1000 ft 1181 pH/ORP, 1050A/1060A,1003/
1023 - 1 mile with standard cable.
4.
Signal cable should be run in a dedicated conduit and should be kept away from AC power
lines.