6
Installation
Installation
Reference Manual
February 2019
00809-0100-3385
4.
When the sensor assembly is properly secured by the ball valve assembly, the valve may
be opened and the sensor positioned into the process at the desired depth and
orientation.
5.
Tighten the hex nut of the male connector to tightly secure the sensor in place during
process conditions. When tightened, the Teflon ferrule inside the hex nut clamps the
sensor
tube. (See Figure 4-3).
The sensor must be captured by the valve assembly and the male connector so that it cannot be blown free
by process pressure if mishandled during insertion or retraction.
CAUTION
Over tightening the hex nut may damage the ferrule.
CAUTION
A stainless steel ferrule is available if the TEFLON ferrule does not inadequately grip. When using the metallic
ferrule, care must be taken to avoid over tightening and damaging the sensor tube. If the male connector
leaks during insertion or retraction, replace the O-ring in the male connector.
NOTICE
Figure 2-1: Typical Mounting Details
Summary of Contents for Rosemount 385
Page 1: ...Rosemount 385 pH ORP Sensors Reference Manual 00809 0100 3385 Rev AA February 2019 ...
Page 2: ...hasgkas ...
Page 4: ......
Page 6: ...Table of Contents Reference Manual February 2019 00809 0100 3385 ii Table of Contents ...
Page 16: ...10 Installation Installation Reference Manual February 2019 00809 0100 3385 ...
Page 26: ...20 Troubleshooting Troubleshooting Reference Manual February 2019 00809 0100 3385 ...
Page 28: ...22 Accessories Accessories Reference Manual February 2019 00809 0100 3385 ...
Page 32: ...26 China RoHS Table China RoHS Table Reference Manual February 2019 00809 0100 3385 ...