1
Introduction
The Rosemount 326T Temperature Transmitter detects media temperature and converts
it into an analog output signal (4 - 20 mA).
1.1
Functions and features
The temperature transmitter converts the measured signal into a temperature-
proportional analog signal. Depending on the parameter setting, the output signal is:
•
4-20 mA with setting OU = I
•
20-4 mA with setting OU = Ineg
The temperature transmitter can scale the analog signal to accommodate the application.
For more technical data, see the Rosemount
™
326T Temperature Transmitter Product
Maximum measuring range when OU = I
Measuring range scaled when OU = I
•
MAW = initial value of the measuring range
•
MEW = final value of the measuring range
•
ASP = analog start point
•
AEP = analog end point
Note
The minimum difference between ASP and AEP is 9 °F (5 °C).
If the temperature is within the measuring range, the output signal is between 4-20 mA. If
the temperature is outside the measuring range, the output signal is between these
ranges:
Temperature condition
Output signal when OU = I
Output signal when OU = Ineg
Temperature > AEP
20-20.6 mA
4-3.7 mA
Temperature > MEW
20.6 mA
3.7 mA
Temperature < ASP
4-3.7 mA
20-20.6 mA
Quick Start Guide
Introduction
00825-0100-4328
January 2019
Quick Start Guide
5
Summary of Contents for Rosemount 326T
Page 1: ...Quick Start Guide 00825 0100 4328 Rev 01 January 2019 Rosemount 326T Temperature Transmitter ...
Page 2: ...2 ...
Page 4: ...Contents Quick Start Guide January 2019 00825 0100 4328 4 Quick Start Guide ...
Page 12: ...Temperature resistance Quick Start Guide January 2019 00825 0100 4328 12 Quick Start Guide ...
Page 14: ...Product certifications Quick Start Guide January 2019 00825 0100 4328 14 Quick Start Guide ...