background image

Quick Start Guide

00825-0100-4104, Rev 02

August 2020

Rosemount

 Hygienic Process

Connection Adapter

Weld-in Adapter

Summary of Contents for Rosemount 326

Page 1: ...Quick Start Guide 00825 0100 4104 Rev 02 August 2020 Rosemount Hygienic Process Connection Adapter Weld in Adapter ...

Page 2: ...ronmental conditions without any restrictions Only use the product for its intended purpose If the operating instructions or the technical data are not adhered to personal injury and or damage to property might occur The manufacturer assumes no liability or warranty for any consequences caused by tampering with the product or incorrect use by the operator Installation electrical connection set up ...

Page 3: ...ling Do not weld the adapter to the container or pipe with the transmitter connected Procedure 1 Bore a hole that has the outside diameter of the adapter in the pipe or container wall Maximum oversize 0 008 in 0 2 mm 2 Weld the adapter Note While welding prevent the adapter from warping a Turn the marking 1 marking may vary to the position shown below for the display of the screwed transmitter 3 ...

Page 4: ...ach other c Apply the welding seams between the fixing points opposite each other Ensure sufficient and even distance between the individual sections to avoid glowing through or warping of the adapter caused by overheating 3 Allow the adapter to cool 4 If used remove the welding mandrel 5 Remove deposits from the thread 4 ...

Page 5: ...m the protective packaging 2 If needed clean the sealing areas 3 Place the O ring in the groove of the transmitter Note Be careful not to scratch or damage the surfaces of the transmitter 326P 1408 326L 327T 4 Grease the transmitter thread with lubricating paste Note The paste must be approved for the application and compatible with the elastomers used 5 ...

Page 6: ...transmitter to the recommended torque Note Over tightening the transmitter can impair the seal If the sealing areas are damaged replace the O ring or the entire process connector 34 mm 39 mm Torque 310 in lb 35 N m 6 ...

Page 7: ...installation meet both the application and 3 A requirements A certificate of compliance is available at Emerson com Rosemount 4 2 Other industry certifications All connector surfaces and materials which come into contact with process medium comply with the following regulations EC No 1935 2004 EC No 2023 2006 CFR Title 21 FDA 177 2600 CFR Title 21 FDA 177 2415 7 ...

Page 8: ...17 USA 1 800 999 9307 or 1 952 906 8888 1 952 949 7001 RMT NA RCCRFQ Emerson com Linkedin com company Emerson Automation Solutions Twitter com Rosemount_News Facebook com Rosemount Youtube com user RosemountMeasurement 2020 Emerson All rights reserved Emerson Terms and Conditions of Sale are available upon request The Emerson logo is a trademark and service mark of Emerson Electric Co Rosemount is...

Reviews: