Transmitter installation
Figure 3-1: Installation Flowchart
Start
Mount the
transmitter
Commissioning
tag
Done
Zero trim the
transmitter
Configuration
Grounding, wiring,
and power up
Set switches and
software write lock
Locate device
3.1
Mount the transmitter
Adjust the transmitter to desired orientation before mounting. Transmitter
must not be securely mounted or clamped in place when changing
transmitter orientation.
3.1.1
Conduit entry orientation
When installing a Rosemount 2051HT, it is recommended installing so a
conduit entry faces downward toward the ground to maximize drainability
when cleaning.
3.1.2
Environmental seal for housing
Thread sealing (PTFE) tape or paste on male threads of conduit is required to
provide a water/dust tight conduit seal and meets requirements of NEMA
®
Type 4X, IP66, IP68, and IP69K. Consult factory if other Ingress Protection
ratings are required.
For M20 threads, install conduit plugs to full thread engagement or until
mechanical resistance is met.
Note
IP69K rating only available on units with a SST housing and option code V9 in
the model string.
Note
For aluminum housings ordered with M20 conduit entries, transmitters
shipped will have NPT threads machined into the housing and a NPT to M20
thread adapter will be provided. Considerations for environmental sealing
listed above should be taken into account when installing the thread
adapter.
November 2019
Quick Start Guide
Quick Start Guide
7
Summary of Contents for Rosemount 2051HT
Page 15: ...Figure 3 6 Configuration Flowchart November 2019 Quick Start Guide Quick Start Guide 15 ...
Page 28: ...Quick Start Guide November 2019 28 Rosemount 2051HT ...
Page 29: ...November 2019 Quick Start Guide Quick Start Guide 29 ...
Page 31: ...November 2019 Quick Start Guide Quick Start Guide 31 ...