background image

7

PENBERTHY

 SERIES LC FLAT GLASS GAGES

INSTALLATION, OPERATION AND MAINTENANCE INSTRUCTIONS

8 REMOVAL - DISASSEMBLY - REASSEMBLY

WARNING

Use only qualified, experienced personnel who 
are familiar with liquid level gage equipment and 
understand the implications of the tables and 
all the instructions thoroughly. DO NOT proceed 
with any maintenance unless the liquid level gage 
has been relieved of all pressure or vacuum, has 
been allowed to reach ambient temperature and 
has been drained or purged of all fluids. Failure 
to follow instructions can cause serious personal 
injury and property damage.

7.2 Troubleshooting

Problem:  Glass becomes etched or clouded 

prematurely in service.

Cause: 

Fluid being handled is not 
compatible with the glass or shields.

Solution:  Replace the glass and install 

shields which will not be affected 
by contained fluid.

Problem:  Glass breaks continually in 

service despite careful attention 
to maintenance procedures.

Cause: 

Thermal shock, hydraulic shock, 
mechanical loads, exceeding design 
ratings or a combination of these.

Solution:  Check entire system to determine 

possible sources of loads. Check 
application to determine actual 
operating conditions and contact an 
authorized Penberthy distributor on 
how to proceed.

FIGURE 3

FIGURE 4
Nut loosening sequence

8.1 Disassembly

Secure a workbench, equipped with a vise at 
one end, longer than the liquid level gage and 
sufficiently wide to lay out parts as they are 
removed. Attach and tighten a nipple to each 
end connection to serve as temporary means 
of supporting the liquid level gage during 
disassembly and reassembly.
1.  Grip nipple on one end with vise and hold up 

the other end with a 2 x 4 or suitable prop as 
shown in Figure 3.

2.  Hold gage firmly; loosen nuts starting 

at both ends of each section and then 
proceeding from both ends to the center 
of each section as shown in Figure 4.

3.  Nut loosening sequence
  -  Remove nuts, washer, belleville washers 

(if any) and nameplate.

  -  Tap covers with rubber hammer as needed 

to loosen and remove.

  -  For belleville washer assemblies: 

to remove covers, studs may need to be 
removed by laying the assembly on its side 
and knocking the stud/U-bolts through the 
cover with a hammer and punch.

  -  Remove cushions, glass, shields (if any) 

and gaskets.

  -  Tap liquid chamber or remaining covers as 

necessary with rubber hammer to break 
loose and remove remaining components.

  -  Remove, destroy and dispose of all glass, 

cushions, gaskets and shields. Under no 
circumstances should these components 
be re-used or installed on a gage.

Note: 

if size of gage is smaller than shown, 

follow spiraling sequence from the ends until 
all bolting is loosened.

WARNING

Once used, cushions, gaskets and shields are 
deformed permanently by compression and, if 
re-used, may cause leaks and high stress points 
resulting in glass breakage. Glass may contain 
hidden damage and internal stresses caused by 
previous usage. If re-used, the glass may break 
under pressure causing severe personal injury 
or property damage.

Vise

Prop

Connection leaks

 at a flanged or threaded 

connection should be corrected by tightening 
the bolting at the connection or by taking the 
liquid level gage out of service and wrapping 
the connection threads with PTFE tape on all 
male pipe threads.

Corrosion

 may occur if the user has selected 

an improper material for the liquid level gage 
application. It is the responsibility of the user to 
choose a material of construction compatible 
with both the contained fluid and the 
surrounding environment. If internal or external 
corrosion is present, an investigation must be 
performed by the user immediately. It may be 
necessary to contact an authorized Penberthy 
distributor to better determine the origin of the 
corrosion.

Summary of Contents for PENBERTHY LC Series

Page 1: ...E INSTRUCTIONS TABLE OF CONTENTS Product warranty 1 1 About the manual 2 2 Introduction 2 2 1 System description 2 3 Available models 3 3 1 Design ratings at maximum and minimum operating temperatures 3 4 Inspection 4 4 1 Glass inspection 4 4 2 User rating inspection 4 5 Installation 4 5 5 1 Piping strain 4 5 2 Differential thermal expansion 4 5 3 Mirror viewing 4 5 4 Nut retorquing 5 5 5 Bellevil...

Page 2: ...tc Visual indication can be enhanced by using reflex glass or illuminators accessory 2 1 System description Penberthy Series LC gages comprise six basic components Each component may vary slightly depending on the desired physical and mechanical properties for the gage Use the exploded parts views in Section 11 as additional reference material Chamber provides a pressure retaining metallic channel...

Page 3: ...00 F 38 C 180 psig 1240 kPa at 400 F 204 C Note lower temperatures are possible with metallic material variation e g 316 stainless construction Grafoil gaskets cushions useable to 325 F 198 C The pressure and temperature ratings may deviate from the previous tables if the gasketing materials of construction and or bolting are other than those specified Higher and or lower temperature ratings are a...

Page 4: ...to dust grit tools and any other objects which may scratch chip or break the glass WARNING DO NOT use glass that is chipped or even slightly scratched Glass surface defects weaken the glass which may result in glass breakage and fluid loss under pressure resulting in serious personal injury or property damage 4 2 User rating inspection The user should confirm that 1 The Series LC liquid level gage...

Page 5: ...limit of bolting on a specific gage TABLE 3 BOLT TORQUE VALUES Gasket material ft lb Nm RLC and TLC w Grafoil standard 20 to 25 27 to 34 RLC and TLC w non asbestos optional 35 to 40 48 to 54 Top chem 2000 20 to 25 27 to 34 PTFE and PCTFE Kel F 20 to 25 27 to 34 25 glass filled PTFE 19 to 22 26 to 30 All models with FKM Viton or elastomeric optional 10 to 15 14 to 20 TLC with PCTFE Kel F shields op...

Page 6: ...age during the test Hydrostatically pressure test all installations to at least 100 psig 690 kPa but less than the design pressure and correct any leakage before proceeding Realistic maintenance schedules can only be determined with full knowledge of the services and application situations involved These will be based upon the user s own operating experience with their specific application If bolt...

Page 7: ...th ends of each section and then proceeding from both ends to the center of each section as shown in Figure 4 3 Nut loosening sequence Remove nuts washer belleville washers if any and nameplate Tap covers with rubber hammer as needed to loosen and remove For belleville washer assemblies to remove covers studs may need to be removed by laying the assembly on its side and knocking the stud U bolts t...

Page 8: ...h 0 05 mm tolerance described in the previous section proceed to obtain new glass gaskets cushions and shields if used and proceed to reassemble as follows refer to exploded parts views in Section 11 if needed 1 Clean threads on bolts and nuts to remove all paint rust and scale Apply a light coat of oil to the threads 2 For transparent gages measure distance between bolt heads of each section and ...

Page 9: ...fore any repairs are initiated should the cost exceed the minimum charge If you return a unit under warranty but it is not defective the minimum charge will apply 6 Install shields if used and gasket on glass being careful to keep components centered 7 For transparent style gages align studs to accept liquid chamber and slide liquid chamber into place being careful to keep components centered 8 Fo...

Page 10: ...shion 9 Shield 100 Screw 125 Washer 163 Nameplate 331 Band RECOMMENDED SPARE PARTS Item Description Min qty 100 Screw 2 per sect 4 Nut 2 per sect 48 Glass 1 7 Gasket 2 8 Cushion 2 9 Shield 2 FIGURE 6 RLC reflex FIGURE 7 TLC transparent 11 EXPLODED PARTS DIAGRAMS NOTE Size 9 shown actual gage may be shorter and require fewer bolting requirements Shields are optional ...

Reviews: