CAUTION! Use only genuine Emerson parts and accessories. Failure to comply may
impair the operation of the product and/or void the warranty.
CAUTION! Only operate the product within the framework of a risk-based alarm signaling
concepts.
WARNING! The gas diffusion path can become occluded (moisture, dust, debris, frozen
condensation) over time, resulting in reduced or complete lack of gas detection and
alarming function. Routine visual inspection of the gas detector and bump testing are
recommended to ensure proper gas detection and alarm function. For information on
calibrating the sensor, refer to the “Care and Maintenance” section of this manual.
Safe Connection of Electrical Devices
Introduction • 3
1.4. Safe Connection of Electrical Devices
WARNING! Before connecting this instrument to electrical devices not mentioned in this
manual, consult the manufacturer or a qualified professional. Failure to comply may result
in personal injury and/or damage to the product.