Table 4-5
- Switch Functions
State
Switch 1 Tap
Switch 1 Hold
Switch 2 Tap
Switch 2 Hold
Warm-up
Enable
Bluetooth®
Connectivity
Disable
Bluetooth®
Connectivity
Normal
Start Zero
Calibration
Start Span Calibration
Low Alarm
Mute Buzzer
Ack. Latched Alarm
High Alarm
Mute Buzzer
Ack. Latched Alarm
Offline
Fault
Mute Buzzer
Ack. Latched Fault
Negative Gas
Fault
Mute Buzzer
Start Zero Calibration
Zero Cal. Fault
Acknowledge
Fault
Span Cal. Fault
Acknowledge Fault
MRLDS-400 Series Application
Operation • 17
4.1.6. Reset System to Factory Default Settings
To reset system to factory defaults, remove lid and hold
TACT#1
and
TACT#2
simultaneously for 30 seconds. Instrument will restart to confirm factory reset.
Alternatively, see
Section 4.2.3.4., Reset to Factory Defaults
, for instructions on
resetting instrument configuration via the MRLDS-400 Series App.
4.2. MRLDS-400 Series Application
Download the MRLDS-400 Series App.
The companion smartphone application allows users to perform a variety of
functions to configure and interact with the MRLDS-450 gas detector, including:
• View real-time measurements
• Configure instrument
• Test outputs
• Calibrate/Bump test instrument
• Generate customizable calibration certificates
4.2.1. Enable Bluetooth® Connection
1. Enable Bluetooth® discovery by tapping
MAG#1
for 1-second. (After 10 seconds, the
device will indicate that it is discoverable with audible heartbeat until it has been
paired, discovery has timed-out, or has been canceled.)