2.2.1
Welding the electrode insert to a pipe
Procedure
1.
Cut an aperture in the pipe for the electrode insert part number 24673540B.
(Any debris must be removed from the pipe-work before testing the system).
2.
Fit the electrode insert into the prepared aperture, ensuring that the orientation
slot is pointing downwards.
Take care not to damage the thread or conical seat on the Series III electrode insert.
The cast/lot number on the insert must not be obscured by welding.
90±5°
1.06 in.
(27 mm)
0.04 in. (1 mm)
nominal
Ø1.5 in. (Ø38.1 mm)
minimum
0.13 to 0.63 in.
(3.4 to 16 mm)
Horizontal pipe run
Vertical pipe run