4
Transmitter power and I/O wiring
4.1
Options for wiring
The wiring procedure you follow depends on which electronics option you have.
Table 4-1: Wiring procedures by electronics option
Electronics option
Wiring procedure
Integral transmitter
The transmitter is already connected to the sensor. No wiring is
required between sensor and transmitter. See the transmitter
installation manual for wiring the power and signal cable to the
transmitter.
MVD
™
Direct Connect
™
There is no transmitter to wire. See the
Micro Motion MVD Direct
Connect Meters
manual for wiring the power and signal cable
between the sensor and the direct host.
Integral core processor with
remote transmitter
The core processor is already connected to the sensor. Connect a
4-wire cable between the core processor and transmitter. Refer
to
.
Remote core processor
attached to transmitter
Connect a 9-wire cable between the sensor and the transmitter/
core processor. Refer to
Micro Motion
Micro Motion 9-Wire Flowmeter Cable Preparation
and Installation Guide
.
Remote core processor
separate from transmitter –
double-hop
•
Connect a 4-wire cable between the core processor and
transmitter. Refer to
•
Connect a 9-wire cable between the sensor and the core
processor. Refer to
, as well as the
Micro Motion
Micro Motion 9-Wire Flowmeter Cable Preparation
and Installation Guide
WARNING
Make sure the hazardous area specified on the sensor approval tag is suitable for the
environment in which the sensor will be installed. Failure to comply with the
requirements for intrinsic safety in a hazardous area could result in an explosion
resulting in injury or death.
NOTICE
Fully close and tighten all housing covers and conduit openings. Improperly sealed
housings can expose electronics to moisture, which can cause measurement error or
flowmeter failure. Inspect and grease all gaskets and O-rings.
Installation Manual
Transmitter power and I/O wiring
20002172
February 2020
Installation Manual
17
Summary of Contents for Micro Motion T Series
Page 4: ...Contents Installation Manual February 2020 20002172 iv Micro Motion T Series ...
Page 12: ...Planning Installation Manual February 2020 20002172 12 Micro Motion T Series ...
Page 16: ...Mounting Installation Manual February 2020 20002172 16 Micro Motion T Series ...
Page 24: ...Grounding Installation Manual February 2020 20002172 24 Micro Motion T Series ...
Page 27: ...Installation Manual 20002172 February 2020 Installation Manual 27 ...