Installation Manual
21
Installation
Figure 3-10 Connecting the wires at the core processor
8. Connect power supply wires to the VDC terminals on the core processor (see
). If you are using Micro Motion 4-wire cable, use the red and black wires.
9. Reattach the core processor cover.
CAUTION!
If the core processor is mounted integrally with the sensor, twisting the core processor will
damage the sensor. To avoid damaging the sensor, do not twist the core processor.
CAUTION!
Damaging the RS-485 wires can cause measurement error or meter failure. Damaging the
power supply wires can cause meter failure. When replacing the core processor cover, make
sure that the wires are not caught or pinched.
RS-485/B
(Green)
4-wire ground screw
(do not use)
VDC –
(Black)
VDC +
(Red)
RS-485/A
(White)
RS-485/B (Green)
RS-485/A (White)
VDC+ (Red)
VDC– (Black)
Standard core processor
Enhanced core processor