10.2
Status alerts, causes, and recommendations
Alert num-
ber
Alert title
Possible cause
Recommended actions
A001
EEPROM Error
The transmitter has detected a
problem communicating with the
sensor.
•
Cycle power to the meter.
•
Replace the core processor.
•
Contact Micro Motion.
A002
RAM Error
The transmitter has detected a
problem communicating with the
sensor.
•
Cycle power to the meter.
•
Replace the core processor.
•
Contact Micro Motion.
A003
No Sensor Response The transmitter is not receiving one
or more basic electrical signals from
the sensor.
•
Check the drive gain and the pickoff
voltage.
•
Check the wiring between the sensor
and the transmitter.
•
Verify that internal wiring is secure and
that there are no internal electrical
problems.
•
Check the integrity of the sensor tubes.
A004
Temperature Over-
range
The RTD resistance is out of range
for the sensor.
•
Check your process conditions against
the values reported by the device.
•
Verify temperature characterization or
calibration parameters.
•
Verify that internal wiring is secure and
that there are no internal electrical
problems.
•
Check the wiring between the sensor
and the transmitter.
•
Contact Micro Motion.
A005
Mass Flow Rate
Overrange
The measured flow rate is greater
than the maximum flow rate of the
sensor (
Δ
T greater than 200 µs).
•
If other alerts are present, resolve those
alert conditions first. If the current alert
persists, continue with the recommen-
ded actions.
•
Check your process conditions against
the values reported by the device.
•
Check for two-phase flow.
A006
Characterization Re-
quired
Calibration factors have not been
entered, or the sensor type is incor-
rect, or the calibration factors are
incorrect for the sensor type.
•
Verify all of the characterization or cali-
bration parameters. See the sensor tag
or the calibration sheet for your meter.
•
Verify the setting of the
Sensor Type
pa-
rameter.
•
If
Sensor Type
=
Curved Tube
, ensure that
no parameters specific to
Straight Tube
have been set.
•
Verify that internal wiring is secure and
that there are no internal electrical
problems.
•
Replace the core processor.
•
Contact Micro Motion.
Troubleshooting
70
Micro Motion
®
LNG Series Meters
Summary of Contents for Micro Motion LNG Series
Page 1: ...Configuration and Use Manual MMI 20032809 Rev AA February 2017 Micro Motion LNG Series Meters ...
Page 6: ...Contents iv Micro Motion LNG Series Meters ...
Page 8: ...Getting Started 2 Micro Motion LNG Series Meters ...
Page 12: ...Before you begin 6 Micro Motion LNG Series Meters ...
Page 18: ...Quick start 12 Micro Motion LNG Series Meters ...
Page 20: ...Configuration and commissioning 14 Micro Motion LNG Series Meters ...
Page 50: ...Configure device options and preferences 44 Micro Motion LNG Series Meters ...
Page 56: ...Integrate the meter with the control system 50 Micro Motion LNG Series Meters ...
Page 58: ...Complete the configuration 52 Micro Motion LNG Series Meters ...
Page 60: ...Operations maintenance and troubleshooting 54 Micro Motion LNG Series Meters ...
Page 74: ...Measurement support 68 Micro Motion LNG Series Meters ...
Page 94: ...Using ProLink III with the transmitter 88 Micro Motion LNG Series Meters ...
Page 98: ...Default values and ranges 92 Micro Motion LNG Series Meters ...
Page 99: ...Default values and ranges Configuration and Use Manual 93 ...