Micro Motion 7827 Digital Viscosity Meter
5
Performance
Temperature specification
Integral temperature sensor
Viscosity Calibrated Range
0.5 to 12,500 cP
Maximum Viscosity Operating Range
0.5 to 20,000 cP (using up to four calibrated ranges)
Viscosity Accuracy
±1% full scale of the operating calibrated range
Viscosity Repeatability
±0.5% of reading
Density Calibrated Range
0.65 to 1.25 g/cc
38 to 78 lb/ft
3
Maximum Density Operating Range
0 to 3 g/cc
0 to 187.28 lb/ft
3
Density Accuracy
±0.001 g/cc of reading
±0.0624 lb/ft
3
Density Repeatability
±0.0001 g/cc
±0.0062 lb/ft
3
Process Temperature Effect
(Corrected)
(1)
(1) Temperature effect is the maximum measurement offset due to process fluid temperature changing away from the factory calibration
temperature.
0.0001 g/cc (per °C)
0.1 k/gm
3
(per °C)
Process
Ambient
Short-stem version
Long-stem version
–40 °F to +185 °F (–40 °C to +85 °C)
–58 °F to +392 °F (–50 °C to +200 °C)
–40 °F to +302 °F (–40 °C to +150 °C)
Technology
100 Ohms RTD (4 wire)
Accuracy
PT100 BS1904 Class B, DIN 43760 Class B