![Emerson Micro Motion 5700 Installation Manual Download Page 25](http://html.mh-extra.com/html/emerson/micro-motion-5700/micro-motion-5700_installation-manual_104641025.webp)
Figure 4-3: mA Output wiring (internally powered)
A
D
C
B
A. mA Output
B. Channel A, B, or C
C. 820 Ω maximum loop resistance
D. Signal device
4.3.2
Wire the mA Output (externally powered)
Procedure
Wire to the appropriate output terminal and pins.
Figure 4-4: mA Output wiring (externally powered)
A
E
D
C
B
A. mA Output
B. Channel A, B, or C
C. 5–30 VDC (maximum)
D. See
for maximum loop resistance
E. Signal device
Installation Manual
Wiring the channels
MMI-20027478
February 2020
Installation Manual
25