Procedure
Set Density Measurement Unit to the option you want to use.
The default setting for Density Measurement Unit is
g/cm3
(grams per cubic centimeter).
Options for Density Measurement Unit
The transmitter provides a standard set of measurement units for Density Measurement
Unit. Different communications tools may use different labels.
Unit description
Label
ProLink III
Field Communicator
Specific gravity unit
SGU
SGU
Grams per cubic centimeter
g/cm3
g/Cucm
Grams per liter
g/l
g/L
Grams per milliliter
g/ml
g/mL
Kilograms per liter
kg/l
kg/L
Kilograms per cubic meter
kg/m3
kg/Cum
Pounds per U.S. gallon
lbs/Usgal
lb/gal
Pounds per cubic foot
lbs/ft3
lb/Cuft
Pounds per cubic inch
lbs/in3
lb/CuIn
Degrees API
degAPI
degAPI
Short ton per cubic yard
sT/yd3
STon/Cuyd
(1) Non‐standard calculation. This value represents line density divided by the density of water at 60 °F.
4.5.2
Configure two-phase flow parameters
ProLink III
Device Tools > Configuration > Process Measurement > Density
Field Communicator
•
Configure > Manual Setup > Measurements > Density > Slug Low Limit
•
Configure > Manual Setup > Measurements > Density > Slug High Limit
•
Configure > Manual Setup > Measurements > Density > Slug Duration
Overview
The two-phase flow parameters control how the transmitter detects and reports two-
phase flow (gas in a liquid process or liquid in a gas process).
Note
Two-phase flow is also referred to as
slug flow
.
Configure process measurement
Configuration and Use Manual
39
Summary of Contents for Micro Motion 2500
Page 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 237: ...NE 53 history Configuration and Use Manual 229 ...