MAS2600: Intrinsically Safe Installation
Reference Manual
January 2017
Ed 22
31
Installation and Special Considerations for Safe Use
Before installing intrinsically safe units within an approved installation, it is essential to ensure that
there are no deviations from the conditions of the order or the requirements of the approved block
schematic diagram and the following related instructions:
1.
No circuits other than those shown on the drawing are permitted. There must be no intercon-
nection to any other circuits.
2.
The intrinsically safe cabling must not be connected to non-intrinsically safe circuits, through
any plug, socket or other form of electric connector containing the cable.
3.
Adequate precautions must be taken to prevent danger arising from the charging of the intrin-
sically safe circuit by the contact leakage or inductance from any other circuit. To prevent static
electricity do not rub cables or amplifier housing with a dry piece of cloth.
4.
The intrinsically safe circuit must be adequately insulated from earth in the hazardous area and
the insulation of all cabling must be at least 500 V r.m.s.
5.
The safety barrier must be placed in a safe area and be mounted within a protective enclosure
with a protection degree, at least equal to IP 20.
6.
The cable between the safety barrier and the amplifier should be provided with a cable shield
which is connected to earth at the safety barrier. The impedance between the safety barrier
earth rail and a secure bond to the hull must not exceed 1 ohm.
7.
Safe area apparatus must not be supplied from, nor contain, a source of potential with respect
to earth under normal or abnormal conditions exceeding 250V r.m.s. or 250 V DC.
8.
The inductance/resistance ratio, total inductance and total capacitance, of the cable and the
amplifier must not exceed the values stated by the manufacturer of the approved safety barri-
ers.
9.
Prevent impacts on the sensor housing to avoid sparks. Do not mount or dismount the sensor
when there is a hazardous condition in the tank. Be sure the tank is vented and gas free.
10.
Be sure that the amplifier box is dry and clean before the lid is mounted. The four screws in the
lid should be tightened fully.
11.
Be sure to order and use sensor cable with the right temperature specification. Check that the
ambient temperature for the entire cable run do not exceed the max limit for the cable.
12.
MAS 2600 contain no repairable parts. Repairing or fixing the circuit or replacing components
may impact the intrinsic safety.
13.
The sensor cable shall remain in one piece from sensor to amplifier box.
Reference is made to manufacturer instructions of safety barriers and to the rules in force at the time
in question concerning intrinsically safe area installation.
Summary of Contents for MAS 2600
Page 1: ...Reference Manual Ed 22 January 2017 MAS 2600 Tank Contents Transmitter ...
Page 4: ...Reference Manual MAS2600 Table of Contents Ed 22 January 2017 2 ...
Page 6: ...Reference Manual MAS2600 Description and Technical Data Ed 22 January 2017 4 ...
Page 10: ...Reference Manual MAS2600 Description and Technical Data Ed 22 January 2017 8 ...
Page 11: ...MAS2600 Ordering Information Reference Manual January 2017 Ed 22 9 Ordering Information ...
Page 28: ...Reference Manual MAS2600 Ordering Information Ed 22 January 2017 26 ...
Page 30: ...Reference Manual MAS2600 Intrinsically Safe Installation Ed 22 January 2017 28 ...
Page 38: ...Reference Manual MAS2600 Installation Ed 22 January 2017 36 ...
Page 44: ...Reference Manual MAS2600 Testing and Recalibration Ed 22 January 2017 42 ...
Page 50: ...Reference Manual MAS2600 Index Ed 22 January 2017 48 ...
Page 51: ......