MODEL Xmt-P

INSTALLATION

5

UNPACKING AND INSPECTION

Inspect the shipping container. If it is damaged, contact
the shipper immediately for instructions. Save the box. If
there is no apparent damage, unpack the container. Be
sure all items shown on the packing list are present. If
items are missing, notify Emerson Process Management
immediately.

INSTALLATION

1. Although the transmitter is suitable for outdoor use,

do not install it in direct sunlight or in areas of extreme
temperatures.

2. Install the transmitter in an area where vibrations and

electromagnetic and radio frequency interference are
minimized or absent.

3. Keep the transmitter and sensor wiring at least one

foot from high voltage conductors. Be sure there is
easy access to the transmitter. 

4. The transmitter is suitable for panel (Figure 3), pipe

(Figure 4), or surface (Figure 5) mounting.

5. The transmitter case has two 1/2-inch (PG13.5) con-

duit openings and either one or four 1/2-inch knock-
outs. The panel mount Xmt-P-FF/FI has four knock-
outs. The pipe/surface mount transmitter has two
knockouts*. One conduit opening is for the power/out-
put cable; the other opening is for the sensor cable. 

Figure 1 shows how to remove a knockout. The
knockout grooves are on the outside of the case.
Place the screwdriver blade on the inside of the case
and align it approximately along the groove. Rap the
screwdriver sharply with a hammer until the groove
cracks. Move the screwdriver to an uncracked portion
of the groove and continue the process until the
knockout falls out. Use a small knife to remove the
flash from the inside of the hole.

6. Use weathertight cable glands to keep moisture out to

the transmitter. If conduit is used, plug and seal the
connections at the transmitter housing to prevent
moisture from getting inside the instrument. 

7. To reduce the likelihood of stress on wiring connec-

tions, do not remove the hinged front panel (-11 mod-
els)  from the base during wiring installation. Allow
sufficient wire leads to avoid stress on conductors.

*NEMA plug may be supplied instead of knockout for
pipe/surface version.

FIGURE 1.  Removing the Knockouts

FIGURE 2.  Power Supply/Current Loop Wiring

(9 - 32 VDC)

Summary of Contents for FOUNDATION Solu Comp Xmt-P-FF

Page 1: ... have received the correct model and options from your purchase order Verify that this man ual covers your model and options If not call 1 800 854 8257 or 949 757 8500 to request correct manual For clarification of instructions contact your Rosemount representative Follow all warnings cautions and instructions marked on and supplied with the product Use only qualified personnel to install operate ...

Page 2: ...s a stabilization check on the sensor in each buffer A manual two point calibration is made by immersing the sensor in two different buffer solutions and entering the pH values The microprocessor automatically calculates the slope which is used for self diagnostics An error message will be displayed if the pH sensor is faulty This slope can be read on the display and or manually adjusted if desire...

Page 3: ...y 381 pHE 31 41 52 Glass only 381 Glass and Reference 385 Glass and Reference 389 02 54 389VP 54 Glass only 396 54 62 396VP Glass only 396P 55 396PVP 55 Glass and Reference 396R 396RVP 54 Glass and Reference 397 54 62 Glass only 398 54 62 398VP 54 Glass only 398R 54 62 398RVP 54 Glass only 399 09 62 399VP 09 399 14 Glass only Hx338 Glass only Hx348 Glass only TF396 none ...

Page 4: ...Xm mt tr r Sensor JBox Temperature in C C F 1 Refer to page 5 for installation instructions 2 Wire pH or ORP sensor to the transmitter See Figure 6 for panel mount Figure 7 for pipe or surface mount Refer to the sensor instruction sheet for details 3 Once connections are secure and verified apply power to the transmitter 4 When the transmitter is powered up for the first time Quick Start screens a...

Page 5: ...F FI has four knock outs The pipe surface mount transmitter has two knockouts One conduit opening is for the power out put cable the other opening is for the sensor cable Figure 1 shows how to remove a knockout The knockout grooves are on the outside of the case Place the screwdriver blade on the inside of the case and align it approximately along the groove Rap the screwdriver sharply with a hamm...

Page 6: ...MODEL Xmt P INSTALLATION 6 FIGURE 3 Panel Mount Installation Access to the wiring terminals is through the rear cover Four screws hold the cover in place Panel Mounting MILLIMETER INCH ...

Page 7: ...MODEL Xmt P INSTALLATION 7 FIGURE 4 Pipe Mount Installation The front panel is hinged at the bottom The panel swings down for access to the wiring terminals Pipe Mounting MILLIMETER INCH ...

Page 8: ...8 MODEL Xmt P INSTALLATION FIGURE 5 Surface Mount Installation The front panel is hinged at the bottom The panel swings down for access to the wiring terminals Surface Mounting MILLIMETER INCH ...

Page 9: ...9 MODEL Xmt P INSTALLATION FIGURE 6 Loop Power and Sensor Wiring Panel Mount ...

Page 10: ...10 FIGURE 7 Loop Power and Sensor Wiring Pipe Surface Mount ...

Page 11: ...MAIN MENU MENU TREE FOR MODEL SOLU COMP Xmt P FF FI TRANSMITTER 11 ...

Page 12: ...N LINE ORDERING NOW AVAILABLE ON OUR WEB SITE http www raihome com Specifications subject to change without notice Emerson Process Management Liquid Division 2400 Barranca Parkway Irvine CA 92606 USA Tel 949 757 8500 Fax 949 474 7250 http www raihome com Rosemount Analytical Inc 2006 ...

Reviews: