![Emerson Fisher 4660 Instruction Manual Download Page 17](http://html.mh-extra.com/html/emerson/fisher-4660/fisher-4660_instruction-manual_102283017.webp)
Instruction Manual
D200163X012
4660 Pressure Pilot
June 2017
17
1. Remove the case cover screw (key 5) and the case cover (key 3).
2. Lift the end of the flapper off the nozzle beam approximately 19 mm (0.75 inch) by moving the opposite end of the
flapper clockwise for a slight rotation.
CAUTION
Be careful not to force the flapper such that the spring (key 18) will become damaged.
3. Release the upper end of the flapper such that the flapper will snap back and contact the nozzle beam.
4. Replace the case cover, and attach it with the case cover screw.
Startup
Verify that the desired high and/or low set point pressures have been set. Refer to the Mounting section for the desired
pilot installation procedures.
Reduce the external process pressure source to zero and disconnect the source from the process
Performance
The performance characteristics shown in figure 6 illustrate several important functional parameters.
Figure 6. Performance Characteristics
TRIP‐TO‐
RESET ZONE
SET POINT
Δ
P
MIN
100%
OUTPUT
PRESSURE
PROCESS PRESSURE
A2897‐2
CONSTRUCTION
Set Point
D
P
MIN
PERCENT OF BOURDON TUBE RATING
Trip‐to‐
Reset Zone
Repeatability
Single High‐Low Unit
High‐Only/Low‐Only Pair
10
(1)
3
≤
1.5
≤
0.25
1. 5000 and 7500 psig Bourdon tubes are 15 percent of Bourdon tube rating.
S
L
and S
H
represent the low and high set points, respectively. The set range is 3 to 97 percent of the Bourdon tube
rating. However, with a single high‐low unit or a high‐only/low‐only pair, there is a limit on how close to each other the
set points can be adjusted. This limit is defined as set point
D
P
MIN
and is shown as S
L'
and S
H'
.
Trip‐to‐reset is the combined effect of pilot dead band and hysteresis. After the pilot has tripped, it will automatically
reset when the process pressure returns to the set range. However, full output is not instantaneous. The difference
between the set point and reset to full output is the trip‐to‐reset zone. This parameter is also a function of the Bourdon
tube rating as shown in figure 6.
Finally, repeatability is the switch point deviation around the set point as a percentage of the Bourdon tube rating.