Under the measurement interface, press and hold [Hold] key (> 3 seconds) to enter the
menu setting interface, adjust the setting contents by [Up] key and [Down] key, then
press [Hold] key (<1 second) ) and select the menu mode. After adjusting the menu to
be modified by [Up] and [Down] key, press [Hold] key (<1 seconds) to save and exit the
menu.
5.4 Sample test
Please follow the points below to get the most accurate results in your measurements:
1. Turn off the induction frying pan during the measurement because the
electromagnetic field will affect the measurement results.
2. Please clean the probe before each measurement or before the next continuous
measurement. See the cleaning instructions of probe for details.
3. Try to avoid touching the probe with metal objects. For example, frying baskets, pot
walls, because these things may affect the measurement results, the minimum spacing
from metal objects is at least 1cm.
4. Uneven oil temperature in the frying oil may cause measurement errors. Please stir the
instrument quickly in the frying oil.
5. If the measurement result is suspected to be wrong because it contains water: Please
repeat the measurement after 5 minutes (do not fry during this period to keep the grease
high temperature). If the new reading goes low, measure again after 5 minutes until the
reading is stable.
6. Please replace the frying oil when it reaches 24% TPM. Different countries have
different limit values. Please be sure to replace the frying oil before reaching the limit
value.