Emerson 96-DL User Manual Download Page 4

 

 

ASSEMBLY INSTRUCTIONS 

 

 

 

 

 

 

 

 

 

- 3 - 

 

1.  Please check to ensure that the package you have received contains the following: 
 

1 each Base containing:  

1 each Hardware Package containing:  

4 each casters 

1 each 5/16 - 18 x 1" hex head screws 

4 each 5/16"

 

lock washers 

1 each 5/16" split lock washer 

1 each Pole Assembly 

 

 

 

1 each 5/16" flat washer 

1 each Head Assembly 

4 each #8 - 32 x 1/2" button head screws 

2 each Bulbs 

4 each #8 lock washers 

 

1 each 3/32" Allen wrench 

 
2.  Remove the Base from it's box and unwrap it from it's padding. Screw the four casters into the base using the 5/16" lock 

washers. Unwrap the Pole Assembly and position in the Base Assembly as shown in Fig. 1. Secure it in place using a 5/16 - 18 
screw, a 5/16" lock washer and a 5/16" flat washer. 

 

3.  Remove the Head Assembly from the package and unwrap it. Place it, bezel side down, on a work surface with the bar facing 

you. While holding the lamps, rotate the Tee Bar up approximately 45 degrees. Lean the Base/Pole Assembly forward so that 
the mount plate at the top of the pole meets the center of the Tee Bar. Connect the cable from the pole to the cable from the bar 
and push the cable and any excess wiring into the bar. Bring the mount plate and Tee Bar together and secure using the four 
8-32 x 1/2" button head screws and four #8 lock washers. 

 

4.  Stand the Light Assembly upright. Install the two bulbs following the "Bulb Replacement" instructions on page 6.  
 
 

OPERATING INSTRUCTIONS 

 

Initial Setup 

 

Plug the power cord into a suitable grounded electrical outlet (110-120 volts, 60 Hz, unless otherwise specified). 

 

Light Position 

 

Position the light over the patient. Grasp the loop on the measuring tape, located on the right lamp assembly, and extend the tape to 
the double red lines, which are 28" (71 cm), from the end. With these lines aligned with the bottom of the lamps and the loop tip 
just touching the patient, you will attain the spacing of 28" (71 cm) between the lamps and the patient that clinical experience has 
shown is suitable for maintaining the patient's temperature during limited periods. 

 
Height Adjustment 

 

Hold onto the upper pole while pulling the height adjustment knob out. Move the upper pole to the desired height and release the 
knob. The spring-loaded knob should engage one of the holes in the upper pole, locking it in position. Once the proper distance has 
been achieved, attach the safety pin in the next higher hole. 

 

Lamp Adjustment 

 

Grasping the bezel handle and rotating the lamp assembly easily changes each lamp. Stops are provided to limit the rotation. 

 

 
 
 
 
 
 
 

 

Summary of Contents for 96-DL

Page 1: ...Medical Product Innovations Since 1928 Models 96 SL 96 DL 96 TDL USER S GUIDE WARMING LIGHTS ...

Page 2: ...m well lit environment is provided for the patient while the room remains comfortable to personnel Procedures may be performed without interference to the thermal environment of the patient A measuring tape is included in the system so that the desirable height adjustment can be made quickly and easily Emerson Warming Lights have been used extensively on both infant and adult patients when mainten...

Page 3: ...w the lights to be closer to the patient than 28 inches 71 cm Monitor the patient s temperature frequently After adjusting the light height ensure that the spring loaded adjustment knob is engaged in one of the holes in the upper pole and the safety pin is inserted in the next higher hole SPECIFICATIONS Model 96 SL Continuously on Model 96 DL Continuously on with adjustable intensity Model 96 TDL ...

Page 4: ... bar Bring the mount plate and Tee Bar together and secure using the four 8 32 x 1 2 button head screws and four 8 lock washers 4 Stand the Light Assembly upright Install the two bulbs following the Bulb Replacement instructions on page 6 OPERATING INSTRUCTIONS Initial Setup Plug the power cord into a suitable grounded electrical outlet 110 120 volts 60 Hz unless otherwise specified Light Position...

Page 5: ...n extended period on infants we suggest shielding the infant s eyes with an opaque material MAINTENANCE Bulb Replacement Loosen the retaining knobs on either side of the lamp housing Gently pull the bezel down and allow it to hang by the retaining wire while you unscrew the bulb After installing a new bulb of the same wattage use Emerson part number 648 3002 push the bezel back on aligning the slo...

Page 6: ...t Lock Washer 4 each Measuring Tape Part No 996 9122 Bezel Retaining Part No 996 9119 Pole Release Knob Lamp Assembly Part No 996 9104 Bezel Screen Assembly Part No 996 9120 Safety Pin Cable Assembly Part No 996 9116 Pole Assembly 96 SL Part No 996 9118 96 DL Part No 996 9139 96 TDL Part No 996 9115 Caster Part No 312 1009 Control Panel Part No 996 9106 96 DL Part No 996 9105 96 TDL End Duct 96 SL...

Page 7: ...Figure 3 Schematic 96 DL Module 6 Figure 4 Schematic 96 DL 96 TDL Module ...

Page 8: ...Figure 5 Schematic 96 DL Module 7 Figure 6 Schematic 96 TDL Module ...

Page 9: ...J H Emerson Co 22 Cottage Park Avenue Cambridge MA 02140 1601 Phone 800 252 1414 U S and Canada or 617 864 1414 Fax 617 868 0841 Email info jhemerson com Web www jhemerson com 910 1109 1 ...

Reviews: