Emerson 96-DL User Manual Download Page 3

 

 

WARNINGS / CAUTIONS / NOTES 

 

 

 

 

 

 

 

 

- 2- 

 

WARNING - ELECTRIC SHOCK HAZARD 

 

Before removing the control panel for servicing, first disconnect the lamp from its electrical power source. 

 

CAUTION 

 

Do not use spray cleaners or disinfectants, or use a wet cloth or sponge to clean the control panel. Liquid can and will enter the 
controls and may cause a malfunction. Instead, clean with a damp cloth. 

 

CAUTION 

 

For prolonged use of lamps on infants, we suggest shielding the patient's eyes with an opaque material. Although there have been 
no reports of injury using the Emerson warming lights, there is concern that retinal or corneal damage may be possible if the patient 
stares continuously at the lamps, due to cumulative exposure to infrared light. When used on adults or for a shorter time, shielding 
the eyes is not necessary. 

 

NOTES 

 

Do not use in hazardous atmospheres. 

 

Use only Emerson part number 648-3002 infrared-type, reflective backed, lightly frosted, uncolored, bulbs. 

 

Do not allow the lights to be closer to the patient than 28 inches (71 cm). Monitor the patient's temperature frequently. 

 

After adjusting the light height, ensure that the spring-loaded adjustment knob is engaged in one of the holes in the upper pole and 
the safety pin is inserted in the next higher hole. 

 

SPECIFICATIONS 
 

Model 96-SL: 

Continuously on 

Model 96-DL: 

Continuously on with adjustable intensity 

Model 96-TDL: 

Timer/Dimmer with a 15-minute adjustable timer and adjustable intensity 

Upper Pole: 

Aluminum pole with detents every 1-1/2" 

Base:   

4 casters, 2" diameter, approximately 20" x 25" 

Power cord: 

10 foot, 16-3 SJTO cord with grounded hospital grade plug 

Overall height: 

Down position: 63"; Up position: 83" 

Lamp Height: 

Down position: 53"; Up position: 73" 

Weight: 34 

lbs. 

Shipping Weight: 

40 lbs. 

Power Requirements: 

110 volts, 60 Hz, 500 Watts (a 220 volt model is available) 

 

DESCRIPTION OF INDICATORS AND CONTROLS 

 

On/Off Switch: 

2-position rocker switch. 

Circuit Breaker Button: 

Safety fuse in the event of electrical malfunction. 

Time Adjustment Control: 

Variable control from 1 to 15 minutes. 

Timer Start Button: 

Push-button switch. 

Dimmer Control: 

Variable control to adjust heat intensity. 

 
 
 
 
 
 

Summary of Contents for 96-DL

Page 1: ...Medical Product Innovations Since 1928 Models 96 SL 96 DL 96 TDL USER S GUIDE WARMING LIGHTS ...

Page 2: ...m well lit environment is provided for the patient while the room remains comfortable to personnel Procedures may be performed without interference to the thermal environment of the patient A measuring tape is included in the system so that the desirable height adjustment can be made quickly and easily Emerson Warming Lights have been used extensively on both infant and adult patients when mainten...

Page 3: ...w the lights to be closer to the patient than 28 inches 71 cm Monitor the patient s temperature frequently After adjusting the light height ensure that the spring loaded adjustment knob is engaged in one of the holes in the upper pole and the safety pin is inserted in the next higher hole SPECIFICATIONS Model 96 SL Continuously on Model 96 DL Continuously on with adjustable intensity Model 96 TDL ...

Page 4: ... bar Bring the mount plate and Tee Bar together and secure using the four 8 32 x 1 2 button head screws and four 8 lock washers 4 Stand the Light Assembly upright Install the two bulbs following the Bulb Replacement instructions on page 6 OPERATING INSTRUCTIONS Initial Setup Plug the power cord into a suitable grounded electrical outlet 110 120 volts 60 Hz unless otherwise specified Light Position...

Page 5: ...n extended period on infants we suggest shielding the infant s eyes with an opaque material MAINTENANCE Bulb Replacement Loosen the retaining knobs on either side of the lamp housing Gently pull the bezel down and allow it to hang by the retaining wire while you unscrew the bulb After installing a new bulb of the same wattage use Emerson part number 648 3002 push the bezel back on aligning the slo...

Page 6: ...t Lock Washer 4 each Measuring Tape Part No 996 9122 Bezel Retaining Part No 996 9119 Pole Release Knob Lamp Assembly Part No 996 9104 Bezel Screen Assembly Part No 996 9120 Safety Pin Cable Assembly Part No 996 9116 Pole Assembly 96 SL Part No 996 9118 96 DL Part No 996 9139 96 TDL Part No 996 9115 Caster Part No 312 1009 Control Panel Part No 996 9106 96 DL Part No 996 9105 96 TDL End Duct 96 SL...

Page 7: ...Figure 3 Schematic 96 DL Module 6 Figure 4 Schematic 96 DL 96 TDL Module ...

Page 8: ...Figure 5 Schematic 96 DL Module 7 Figure 6 Schematic 96 TDL Module ...

Page 9: ...J H Emerson Co 22 Cottage Park Avenue Cambridge MA 02140 1601 Phone 800 252 1414 U S and Canada or 617 864 1414 Fax 617 868 0841 Email info jhemerson com Web www jhemerson com 910 1109 1 ...

Reviews: