34
“Options” menu.
Press wheel and rotate to choose.
TAU: if probes reading values are displayed too fast
increase TAU value to stabilize them. Default value is 03.
Maximum value is 30.
Coeff Temp: Temperature compensation. Minimum value
is 0% (disabled), maximum value is 5%. See page 15 for
more info.
Startup Delay: Startup time delay at instrument startup
(power on). Minimum value is 0m (disabled), maximum
value is 99minutes.
FLOW: Flow contact (34-35-36) working mode.
Set to “NORMAL” for Normally Open (N.O.) contact type.
Set to “REVERSE” for Normally closed (N.C.) contact type.
Set to “DISABLED” to disable contact.
OUT CURRENT: mA current output configuration.
Set to 4/20 mA for 4 (setpoint low) and 20 (setpoint high) mode.
Set to 0/20 mA for 0 (setpoint low) and 20 (setpoint high) mode.
Cd mA High and Cd mA Low are related
to maximum and minimum conductivity value
for “mA output 1”.
Temp mA HI and Temp mA LO
are related
to maximum and minimum temperature value
for “mA output 2”.
pH mA High and ORP mA Low are related
to maximum and minimum pH value
for “mA output 3”.
mV mA High and ORP mA Low are related
to maximum and minimum ORP value
for “mA output 4”.
NUMBER WEEK: Set weeks number for “Biocide 1” and
“Biocide 2” working mode. See page 18 for more info.
FACT. DEFAULT: it reverts instrument to factory settings
including PASSCODE (“0000”).
IF ALARM: set this option to DISABLE if mA outputs must
be disabled during an alarm event or ENABLE to let mA
outputs enabled during an alarm event.