![Emdamed Non-Contact Forehead InfraRed Thermometer User Manual Download Page 9](http://html1.mh-extra.com/html/emdamed/non-contact-forehead-infrared-thermometer/non-contact-forehead-infrared-thermometer_user-manual_2399949009.webp)
Non-Contact Forehead IR Thermometer User Manual
Non-Contact Forehead IR Thermometer User Manual
17
16
This message displays in various cases - please find
belo of the main cases.
15. This appliance conforms to the following standards:
•
ASTM E 19651998
•
EN 980: Graphical symbols for use in the labeling of medical
devices
•
EN 1041:Information supplied by the manufacturer with
medical devices
•
EN 60601-1: Medical electrical equipment Part 1: General
requirements for safety (IEC:60601-1:1998)
•
EN 60601-1-2: Medical electrical equipment Part 1-2: Gener-
al requirements for safety Collateral standard Electromag-
netic compatibility Requirements and test (IEC 60601-1-
2:2001)
EMC Statement
This device has been tested and homologated in accordance
with EN 60601-1-2:2007 for EMC. This does not guarantee in
any way that the device will not be affected by electromagnetic
interference. Avoid using the device in high electromagnetic
environment.
The MEDICAL ELECTRICAL EQUIPMENT needs special precau-
tions regarding EMC and needs to be installed and put into ser-
vice according to the EMC information provided in the ACCOM-
PANYING DOCUMENTS.
Reasons for “LO” or “HI”
message display
Advice
Temperature reading ham-
pered by hair, perspiration...
Make sure that there is no
obstruction prior to taking a
temperature.
Temperature hampered by
an air flux
Make sure there is no air flux
as this could interfere with
the infrared system.
The measuring distance is
too far.
Please respect the measur-
ing distance (between 5 to
15cm - 2 to 5.9 in).
From high/low temperature
condition to room temper-
ature
Waiting for 10 minutes
before taking the body tem-
perature
Summary of Contents for Non-Contact Forehead InfraRed Thermometer