Emdamed Non-Contact Forehead InfraRed Thermometer User Manual Download Page 8

Non-Contact Forehead IR Thermometer User Manual

Non-Contact Forehead IR Thermometer User Manual

15

14

11-9. Longevity Use 

The Non-contact Body Infrared Thermometer was conceived for 

an intense and professional use, its longevity is a quarantine for 

40000 takings.

12.Advice 

• 

The protective glass over the lens is the most important and 

fragile part of the thermometer, please take great care of it.

• 

Do not recharge non rechargeable batteries, do not throw .

• 

Do not expose the thermometer to sunlight or water.

13. Maintenance and Cleaning

 

• 

The Infrared Sensor is the most precise part, must be pro-

tected carefully

• 

Clean the device with a cotton bud lightly moistened with 

70% alcohol.

• 

Do not clean the device with corrosive detergent.

• 

Keep the device away from water or other liquid. • Store the 

device in a dry environment, and keep it away from dust 

and direct sunlight.

14. Troubleshooting

If you happen to have one of the following problems while using 

your the non-contact forehead IR thermometer please refer to 

this bres service guide to help resolve the problem. If the prob-

lem persists please contact our customer service,

The screen displays the body temperature inferior to 32°C 

(89.6°F) If you’re on Surface mode the 32°C(89.6°F) temperature 

displayed is showing the external temperature that your body 

releases.

The screen displays the message “HI”

When using the Non-contact Body Infrared Thermometer the 

message HI can show on the screen.

The analysis is above the measurement range selected, either 

superior to 42.5°C (108°F) in Body mode or superior to 60°C 

(140°F) in Surface mode.

The screen displays the message “LO”

When using the Non-contact Body Infrared Thermometer the 

message Lo can show on the screen.

The temperature analyzed is under the measuring range select-

ed, either less than 32°C (90°F) in Body mode or less than 0°C 

(32°F) in Surface mode.

Summary of Contents for Non-Contact Forehead InfraRed Thermometer

Page 1: ...eter User Manual Please read this manual carefully or purchase and use it under the guidance of the medical staff For prohibited content or precautions please refer to this manual Please keep the manu...

Page 2: ...4 2 Safety Information 4 3 Features 5 4 Intended Use 6 5 Configuration 7 6 Indicator 7 7 Description of Symbols 8 8 Technical Specifications 9 9 Calibration Instructions 10 10 Use 11 11 Measuring Oper...

Page 3: ...erspiration or other factors please take the temperature behind the ear lobe See Part 10 5 Clean the glass with a cotton bud lightly moistened with 70 alcohol Importance Before taking of the temperatu...

Page 4: ...ype and thickness of skin Normal Temperatures According To Measurement Method Measurement Method Normal Temp C Normal Temp F Rectal 36 6 to 38 97 8 to 100 4 Oral 35 5 to 37 5 95 9 to 99 5 Axillary 34...

Page 5: ...tive symbol Follow the instruction for use Keep dry Manufacturer symbol 7 Description of Symbols 8 Technical Specifications Normal Conditions of Use Display Resolution 0 1 C 0 1 F Operating Temperatur...

Page 6: ...i e 1 1 C 2 2 F To do it press the MODE button for 2 seconds the screen displays F1 press MODE button again until you get F3 Press UP button in order to add the difference in our example 1 1 C 2 2 F...

Page 7: ...temperature of a human being For this use the Body mode Measurement range for Body mode 32 to 42 5 C 86 to 108 F You can also use the Non contact Body Infrared Thermometer to measure the temperature o...

Page 8: ...t away from dust and direct sunlight 14 Troubleshooting If you happen to have one of the following problems while using your the non contact forehead IR thermometer please refer to this bres service g...

Page 9: ...gated in accordance with EN 60601 1 2 2007 for EMC This does not guarantee in any way that the device will not be affected by electromagnetic interference Avoid using the device in high electromagneti...

Page 10: ......