Section 4. Technical Information about easypod
™
Medical electrical equipment needs special p recautions regarding EMC (electromagnetic
compatibility) and needs to be used accord ing to the EMC information provided below.
Portable or mobile RF communications equipment (e.g., mobile phones, pagers…) can
affect easypod.
easypod is in tended for use in the electromagnetic environment specified below. The
user should ensu re that easypod is used in such an environment.
Guidance and manufactu rer’s declaration – electromagnetic emission
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions CISPR 11
Group 1
easypod uses RF energy only for its internal function. Therefore, its RF emissions
are very low and not likely to cause interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations / flicker
emissions IEC 61000-3-3
Not applicable
easypod is suitable for use in all establishments, including domestic establishments
and those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Guidance and manufacturer’s declaration – electromagnetic immunity
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment – guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrostatic transient / burst
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines
Not applicable
Not applicable
Surge IEC 61000-4-5
± 1 kV differential mode
± 2 kV common mode
Not applicable
Not applicable
Voltage dips, short interruptions
and voltage variations on power
supply input lines IEC 61000-4-11
Not applicable
Not applicable
Not applicable
Power frequency (50/60 Hz)
Not applicable
Not applicable
Not applicable
Electromagnetic Compatibility Tables