4 Perform a test

4.5.2 EUT-Test

1

2

3

4

5

1 Generator

Select the desired Waveform and Ground Injection as coupling type.

2 Connection cable

The 2m MC-cable is connected between the EUT and the generator output. The generator is
protected with a 16A / 250V fuse against overcurrent.

3 Voltage monitor

During the test, peak voltage must be monitored to ensure the amplitude does not exceed the test
level. The pulse shape is not relevant.

4 Current monitor

Use a current probe to monitor the current during the test to ensure the peak does not exceed the
limit level.

5 Oscilloscope

The oscilloscope must have a minimum bandwith of 500 MHz.

4.6 Cable Induction WF4

The preferred coupling method for WF4 is Ground Injection. Ground Injection is possible directly
with AVI3000 without any additional coupling device. The CN-GI-CI-V is only neccessary if WF4
will be applied as Cable Induction test.

4.6.1 System Calibration

Prior to each test, the system has to be calibrated. The calibration is made at the desired test level with
both polarities.

The tests can be realized with 1 or 2 turns on the coupler output. With 2 turn, Level 3 can be
achieved with every kind of load. In many cases is it also possible with one turn on the output.
The setting has to be considered in the calibration and later in the EUT test. The ratio is also
selectable in the generator parameters.

Page 24/45

Summary of Contents for AVI3000

Page 1: ...LARGEST RANGE OF IMPULSE TESTERS UP TO 100KV 100KA User Manual Version 1 0 AVI3000 CN BT7 SHUNT0E1 CN GI CI V...

Page 2: ...ARTNER AG fulfill the EMC and safety requirements to obtain CE certification A copy of the individual certificates can be viewed and downloaded from www emc partner com section company documents This...

Page 3: ...perated in a dry room If condensation is visible the affected unit s should be dried before operating Never touch the Equipment under Test EUT when the test system is operating Establish a safety barr...

Page 4: ...n Raum betrieben werden Wenn Kondensation Beschlag sichtbar ist sind die betroffenen Teile der Ausr stung zu trocknen Ber hren Sie niemals die getestete Ausr stung EUT wenn das Testsystem arbeitet Ste...

Page 5: ...s un local sec Si de la condensation est visible les appareils affect s doivent tre s ch s avant utilisation Ne jamais toucher l quipement sous test EST lorsque le syst me est sous fonction Mettre en...

Page 6: ...bration 18 4 3 2 EUT Test 19 4 4 Cable Bundle WF1 WF2 WF3 WF5A WF6 20 4 4 1 Calibration 20 4 4 2 EUT Test 21 4 5 Ground Injection WF4 22 4 5 1 System Calibration 22 4 5 2 EUT Test 24 4 6 Cable Inducti...

Page 7: ...ction 38 10 5 WF3 1MHz Cable Induction 38 10 6 WF3 10MHz Cable Induction 39 10 7 WF4 Pin Injection 39 10 8 WF4 Ground Injection 40 10 9 WF4 Cable Injection 40 10 10WF5A Pin Injection 41 10 11WF5A Cabl...

Page 8: ...comfortable user interface Interactive parameter input help text and graphic illustrations guide the user through the instrument programming process CN BT7 PN 107632 Coupling transformer for Cable Bun...

Page 9: ...round Injection is possible directly with AVI3000 without any additional coupling device The CN GI CI V is only neccessary if WF4 will be applied as Cable Induction test Not all accessories are used f...

Page 10: ...ow can cause the equipment to overheat which may disturb the operation and even cause damage Most equipment from EMC PARTNER can be mounted in a 19 Rack A special rack mounting kit is available Contac...

Page 11: ...erform all functions including test types selection parameter entry saving and recalling tests 3 Rotary knob Entry device enables faster navigation and parameter entry 4 Run Stop Starts or stops the s...

Page 12: ...SB interface is also used to update the internal firmware 7 Auxiliary Connection Interface for additional external equipment 8 Trigger out BNC output for a direct connection to the scope to start puls...

Page 13: ...socket housing The value is indicated on the equipment type plate Take care to ensure the power main matches the power ratings on the equipment type plate The generator must always be connected to pro...

Page 14: ...pressed on the BNC connector is indicated in the test protocol as an EUT Fail condition The input is internally pulled up to max 24V Driving with an open collector output is recommended 7 Trigger Inpu...

Page 15: ...available all the time Availability depends on hardware configuration To simplify the menu structure tests that are unavailable can be removed from the configuration General Menu Settings Test Settin...

Page 16: ...ect a category using the touch panel 2 Parameter list All parameters from a selected category are shown on a single screen Click on a parameter to edit the value Press and turn the rotary knob at the...

Page 17: ...a running test 2 Progress bar Indicates progress of the whole test and if available about the current cycle 3 Status bar Provides information about the current event 4 Running options Click to open a...

Page 18: ...The calibration is performed including a 2m MC cable Connect the cable to the correct output The pictures shows the setup for WF4 and WF5A For WF3 the output is on the front of the plugin The correct...

Page 19: ...Connection cable The 2m MC cable is connected between the EUT and the generator output The pictures shows the setup for WF4 and WF5A For WF3 the output is on the front of the plugin The correct outpu...

Page 20: ...r WF1 connect the WF1 output on the generator with the CN BT7 INPUT 1 using a 2 m HV BNC cable On the coupler put the MC bridge in the FILTER position For WF2 connect the WF2 output on the generator w...

Page 21: ...or WF2 measured directly with SHUNT0E1 on MONITOR1 For WF3 1MHz and 10MHz measured directly with SHUNT0E1 on MONITOR1 For WF5A short circuit MONITOR 2 with a 0 25 m MC cable and measure with a current...

Page 22: ...output on the generator with the CN BT7 INPUT 1 using a 1 m HV BNC cable On the coupler put the MC bridge in the PARKING position For WF5A connect the WF5A output on the generator with the CN BT7 INP...

Page 23: ...cable Connect the cable to the correct output The correct output is always indicated by a red LED 3 Measurement Equipment Use a HV differential probe to measure the open circuit voltage The current is...

Page 24: ...loscope must have a minimum bandwith of 500 MHz 4 6 Cable Induction WF4 The preferred coupling method for WF4 is Ground Injection Ground Injection is possible directly with AVI3000 without any additio...

Page 25: ...generator menu 2 CN GI CI V The output of the generator is connected with a 2m MC cable to the 5 Turn input of the CN GI CI V 3 Measurement Equipment Use a HV Differential probe to measure the open ci...

Page 26: ...of the CN GI CI V The cable bundle of the EUT is wound one or two times through the coupler 3 Voltage monitor During the test the peak voltage must be monitored to ensure the amplitude does not excee...

Page 27: ...opied to a USB memory stick by selecting the file and choosing copy to USB from the menu To copy multiple protocol files select with the check box on the right side When all files to be copied are mar...

Page 28: ...y configured test will be saved 5 2 2 Load setups Choose between the EMCP Library with predefined setups the user defined internal library or setups on the USB stick Select a setup then choose load to...

Page 29: ...t is used for the setup library and the saved protocols 1 2 3 4 5 6 7 1 Root directory Files can be selected from the internal memory or if connected the USB memory stick 2 Selected path Shows the cur...

Page 30: ...can perform different tests from the setup linker During a running test changed parameters can be resetted when the test is over Unavailable tests due to missing hardware can completly be removed ins...

Page 31: ...y Only external accessories available for connection to the test equipment are shown in this list Any test requiring an external accessory will only be available to the user after correct selection ha...

Page 32: ...ion contact your local representative or EMC PARTNER directly V PROBE SI Differential HV Probe up to 7kV with a bandwidth up to 70MHz to measure all Waveforms of AVI3000 during calibration and test I...

Page 33: ...ngle test capability save and recall of files and a comprehensive help file Test routines can be prepared in TEMA3000 and then uploaded into the generator in the test lab Programing with the REPOS int...

Page 34: ...ir filter of the ventilator in soapy water The air filter must be dried before being reinstalled 9 4 Warranty EMC PARTNER s warranty is two years from date of invoice for manufacturing defects and one...

Page 35: ...and select contacts If there is no local organization contact EMC PARTNER headquarters in Switzerland Our contact details are also given on the web site Whenever further assistance is required provid...

Page 36: ...uit Short Circuit WF5A 300V 300A Open Circuit Short Circuit Cable Bundle Single Stroke WF1 600A 300V Test Limit level WF2 300V 600A Test Limit level WF3 600V 120A Test Limit level WF4 300V 600A Test L...

Page 37: ...Stroke Subsequent Strokes Current Risetime 6 4us 20 Duration 69us 20 Voltage Risetime 6 4us Duration 6 4us Coupling Device CN BT7 10 3 WF2 Cable Induction Standard DO 160 Section 22 Coupling Type Cabl...

Page 38: ...ifth Peak Syncro 0 359 10 EUT max Voltage 230V max Frequency 800Hz Coupling Device direct 10 5 WF3 1MHz Cable Induction Standard DO 160 Section 22 MIL STD 461G CS117 Level Single Stroke 50V 1900V Mult...

Page 39: ...mping 25 75 First Peak to Fifth Peak Current Frequency 1MHz 20 Damping 25 75 First Peak to Fifth Peak Coupling Device CN BT7 10 7 WF4 Pin Injection Standard DO 160 Section 22 Voltage 50V 500V 10 0 Cur...

Page 40: ...9us EUT Bypass Current max 16A Coupling Device direct 10 9 WF4 Cable Injection Standard DO 160 Section 22 Level Turn Ratio 5 1 Single Stroke 10V 300V Multiple Stroke 10V 150V 10V 75V First Stroke Subs...

Page 41: ...20 Syncro 90 10 270 10 positive Pulse negative Pulse EUT max Voltage 230V max Frequency 800Hz Coupling Device direct 10 11 WF5A Cable Induction Standard DO 160 Section 22 MIL STD 461G CS117 Level Sin...

Page 42: ...14 Subsequent Transients Duration 1 5s Spacing 10ms to 200ms randomly User defined pattern Pulses 1 First transient up to 30 Subsequent Transients Duration 20ms to 10s Spacing 10ms to 400ms randomly...

Page 43: ...sing Pressure 860hPa 1060hPa System Display 7 Touch 640 x 400px VGA Internal memory 8GB Connections USB Typ A Plug LAN RJ45 10 100 BaseT EXT Trigger Input BNC max 24V EUT Fail Input BNC max 24V Warnin...

Page 44: ...ent 10A 20A 32A WF4 800Hz WF4 400Hz WF4 50 60Hz Turn Ratio 5 2 5 1 up to Level 3 up to Level 2 Dimension 65cm x 53cm x 50cm l x h x w Aperture 6cm x 12cm Weight 190kg 10 18 SHUNT0E1 Dimension 12cm x 2...

Page 45: ...y 9 The product should be recycled through a professional organization with appropriate experience for the disposal and recycling of electronic products EMC PARTNER is also available to help with ques...

Reviews: