2 Safety
The CN16 belongs to safety class 1
2.1 Safety standards
The CN16 fulfils the requirements of the safety standards IEC 61010 „Safety requirements for electrical
equipment for measurement, control and laboratory use and the safety standard VDE 0104 ( Safety circuits,
warning lamps or connector for warning lamps ). Based on EN 61010 (IEC 61010) the declaration of
conformity to low voltage directive LVD 73/23/EEC (O.J. N° L77, 1973-03-26) is given.
This manual is a integral part of the CN16 network. The instructions contained in the manual
regarding operation and the test set up are to be strictly observed.
2.2 Climatic conditions
The CN16 generators contain high voltage circuits in integrated form. EMC PARTNER only guarantees a
correct functioning of the CN16 network and the associated accessories, if the CN16 is operated in the
climatic condition specified.
Temperature
15 °C to 35 °C
60 to 90°F
Relative humidity
45 % to 75 %
12.9 to 15.4 PSI
Atmospheric pressure
86 kPa to 106 kPa
(860 to 1060 mbar)
Not influenced by:
direct solar radiation, rain or condensate water, dust or larger electro
magnetic fields as specified in the EMC compatibility chapter.
The CN16 should be operated in a dry, clean room. If for any reason water condenses in the CN16, then no
CN16 operation should be started before the tester is dry.
It is strictly forbidden to operate the CN16 network in rooms with of gas explosion risk. The high
voltage of the CN16 can generate sparks, which can ignite the gas.
People with heart pacemakers should not be in the vicinity of the test set up during operation.
Summary of Contents for 103538
Page 3: ...Common Mode Coupling Network CN16 3 28 8 SERVICE INFORMATION 23 9 GLOSSARY 25 ...
Page 4: ......
Page 8: ...Common mode coupling network CN16 8 28 ...
Page 20: ...Common mode coupling network CN16 20 28 ...
Page 22: ...Common mode coupling network CN16 22 28 ...
Page 24: ...Common mode coupling network CN16 24 28 ...
Page 27: ...E CN16 E Manual 27 28 ...