4.2.1. Quickstart during instrument calibration
This function needs Firmware 2.11 or higher.
In case of a calibration of the measuring instrument it is necessary to use external signal generators. The CWS
instrument has frequency matched calibration curves for each instrument channel. When using the internal
signal generator CWS 500 knows the actual frequency and select automatically the correct reference. When
using an external signal Generator, the instrument does not know the actual frequency. Therefore it is very
important to adjust the frequency in the display to the actual applied frequency. otherwise the displayed
measuring value can be wrong.
Page 3
(Operating)
Quickstart
AMP :2
Att
= 63,0 dB
f
=
100.000 MHz
Mod = AM 1kHz 80%
FPow = - 3.0 dBm
Mon = -56.9 dBm
Rpow = - 56.5 dBm
-----------------------
f---------------------
STOP
1k
10k
100k
1M 10M
F1
F2
F3
F4
F5
F6
F7
Adjust the frequency when using an external signal generator during instrument calibration