EN
10
Technical Specifications
10.1 BioProTT
TM
Clamp-On SL
Size (HxWxD), Weight (incl. cable & plug)
25 x 33 x 45 mm (IDs 13427; 13423; 13424; 13420; 13421; 13425 & 13426),
137 g
28 x 38 x 51 mm (IDs 13422 & 13428, 13433), 153 g
35 x 43 x 69 mm (IDs 13430; 13429 & 13431), 225-245 g
46 x 56 x 84 mm (IDs 13432; 13434 & 13435), 380-400 g
BioProTT
TM
Clamp-On SL with mounting feature:
Size (HxWxD), Weight (incl. cable & plug)
31 x 33 x 45 mm (ID 13650, 13651); 148 g
34 x 38 x 51 mm (ID 13652); 165 g
Housing and Lid Material
Epoxy Resin, Aluminum, Brass
(for IDs 13427; 13423; 13424; 13420; 13421; 13425; 13426; 13433; 13422 &
13428)
Polymeric Compound, Aluminum, Brass
(for IDs 13430; 13429; 13431; 13432; 13434 & 13435)
Cable length**
2.9 m
±
5 cm
with extension cable: max. 4 m
Connector plug
16-pin round plug
IP-Code
without connector: IP67 (protected against splashing water)
Recommended tube type
flexible, non-reinforced tubing; e.g. silicone, PVC
Possible adjustment temperature range
4 °C to 45 °C (40 °F to 113 °F)
*Please note:
It is possible to adjust the cable length according to customer specifications as long as the total cable length does not
exceed 4 m.
Accuracy* in Combination with the BioProTT
TM
Product Range:
Outer Tubing Diameter (OD)
Accuracy
up to OD 3/4"
±2 % of reading ±20 ml/min
up to OD 7/8"
±2 % of reading ±80 ml/min
up to OD 1-1/8"
±2 % of reading ±200 ml/min
up to OD 1-5/8"
±2 % of reading ±400 ml/min
*The specified accuracies were determined under the following conditions:
Medium: Water
Medium Temperature: Adjustment Temperature
Straight Inlet Section: 15 x inner diameter (ID) of tubing
Fully developed flow profile
Tubing: Tube the sensor was adjusted and calibrated for.
The accuracy stated above can only be guaranteed if the parameters throughout the measurement are the same
as the ones the sensor was adjusted and calibrated for (e.g. medium type, medium temperature, tube size, tube
material).
Please note:
–
The accuracy is specified within the flow measurement range indicated in the table below. The flow measurement
range is limited by the Qmin and the Qmax.
–
Qmin refers to the minimum flow value for which the accuracy is specified.
–
Qmax refers to the maximum flow value for which the accuracy is specified.
–
The given values describe the positive flow range, but are the same for negative flow values.
–
The installation position of the BioProTT
TM
Clamp- On SL as well as the positioning of pumps and valves within
the circuit impact the measurement and must be taken into account when it comes to the accuracy of the
BioProTT
TM
FlowMeasurement System.
BioProTT™ Clamp-On SL
Page 22
Copyright em-tec GmbH
D162-704 BioProTT™ Clamp-On SL – User Manual– V4.0