3.

  After inserting the tube, wait for about 5 minutes before proceeding with the setup to allow the tube 

enough time to regain its original structure. 

4.

  Ensure that the temperature of the medium you use for your application is the same as the one the 

BioProTT

TM

 Clamp-On SL was adjusted and calibrated for. 

5.

  Carry out a zero flow adjustment. For this,

 

allow sufficient time for the system to adapt to ambient conditions.

 

ensure that the sensor is clamped on at the correct position.

 

ensure that there are no air bubbles inside the tube.

 

the tube must be completely filled with liquid, but the liquid 

must not 

move.

 

the acoustic coupling rate must be checked. It should be ≥ 60 %. 

 

A zero flow adjustment must be carried out before each measurement 

to avoid possible offsets 

from the measured values.

During the flow measurement: 

• 

Ensure that the coupling is within the operating range (i.e. ≥ 60% ). If the coupling falls below 50%, depending 
on the respective evaluation device, there is either a warning issued or no flow value is displayed as the  

low coupling might impair the measurement. 

• 

It is important that any unwanted interruption of the process due to damage or kinking of the tubing is 
avoided at all times.

• 

Any changes in the medium itself or its temperature can cause errors or anomalies in the measurement. 

• 

It must be ensured that the correct calibration parameters are used.

• 

While it is possible to carry out measurements outside of the specified flow range, em-tec GmbH cannot 
guarantee that resulting flow values will be within the given accuracy range. 

The parameters of the connected BioProTT

TM

 Clamp-On SL and the selected calibration table are accessible 

via the display or the web interface of the respective BioProTT

TM

 evaluation device.

 

Please note: 

The stated measurement accuracy of the BioProTT

TM

 Clamp-On SL is only valid if the parameters of your 

process and application are the same ones the sensor was calibrated and adjusted for and if your setup 
respects the factors stated in the TechNote "

Optimizing Accuracy

". 

BioProTT™ Clamp-On SL

Copyright em-tec GmbH

Page  13

D162-704 BioProTT™ Clamp-On SL – User Manual– V4.0 

DocuSign Envelope ID: BE88571C-E493-4DFB-AF9A-864EB3FC3D75

Summary of Contents for BioProTT Clamp-On SL

Page 1: ...BioProTTTM Clamp On SL User Manual DocuSign Envelope ID BE88571C E493 4DFB AF9A 864EB3FC3D75...

Page 2: ...c GmbH Finning All rights reserved Subject to Technical Changes Owing to our policy of continuous product development the illustrations and technical data contained in this document may differ slightl...

Page 3: ...Maintenance and Service 9 6 Use of the BioProTTTM Clamp On SL 10 6 1 General Safety Information 10 6 2 Connecting the BioProTTTM Clamp On SL to the Evaluation Device 10 6 3 Attaching the BioProTTTM Cl...

Page 4: ...itoring and control of the upstream feed monitoring and control of the downstream feed filtration fill and finish processes process evaluation upscaling processes research The SL at the end of the pro...

Page 5: ...Clamp On SLs incorporate two pairs of transducers i e four piezo ceramics in total which are arranged in an X configuration around the tube containing the measured flow When sending ultrasonic signals...

Page 6: ...aning Do not dispose of this device together with domestic waste The device as a whole as well as any parts must be disposed of in accordance with WEEE Directive and national legislation Manufacturer...

Page 7: ...Clamp On SL is part of to ensure that the BioProTTTM Clamp On SL is used inspected and maintained in accordance with the user manual This also applies for all subsequent revisions or instructions from...

Page 8: ...8 Troubleshooting as well as to the user manual of the evaluation device respectively 5 5 Electrical Safety and Electromagnetic Compatibility The BioProTTTM FlowMeasurement System i e the BioProTTTM C...

Page 9: ...service for the BioProTTTM Clamp On SL may be carried out by em tec GmbH only If these instructions are not followed em tec GmbH shall accept no liability for the component and the warranty will be vo...

Page 10: ...TTM FlowTrack SL To connect the BioProTTTM Clamp On SL to the BioProTTTM FlowTrack SL follow the steps listed below Carefully attach the round sensor plug to the sensor connection of the evalution dev...

Page 11: ...t deformed damaged or dirty Ensure that the tube size and the tube material used corresponds with the size of the BioProTTTM Clamp On SL Air in the tube can lead to errors in the measurement or interr...

Page 12: ...ease refer to the user manual of the respective BioProTTTM evaluation device Make sure to follow the steps described below and to heed the warnings included in them in order to ensure the highest poss...

Page 13: ...ued or no flow value is displayed as the low coupling might impair the measurement It is important that any unwanted interruption of the process due to damage or kinking of the tubing is avoided at al...

Page 14: ...ction of the BioProTTTM Clamp On SL Make sure that no detergents or disinfectants leak into the sensor plug For the cleaning of the device we recommend commonly used detergents and or the recommended...

Page 15: ...ensor or to fix it in one position em tec offers optional mounting solutions The features differ slightly in regard to their design and handling depending on the sensor size Sensors with an OD of up t...

Page 16: ...or torsial moment caused by the tube are too high the thread may tear the sensor base body may break or the sensor might be levered out If the mounting feature and thus sensor is not positioned corre...

Page 17: ...mmends regular calibrations to be carried out Ideally the time between re calibrations should not exceed 24 months Other than the adjustment the calibration can be carried out by the customer on site...

Page 18: ...nfluence the sensor or if the sensor or its cable came into contact with an electric cautery 2 Check the grounding of the valves and pumps or any other devices that might create electromagnetic distur...

Page 19: ...nsor lid is closed properly 2 Close the lid If the problem persists contact our service department The tube size and or material is not the one indicated for the sensor 1 Check if the tube size and ma...

Page 20: ...hannel If the problem persists contact our service department There might be electromagnetic fields influencing the sensor 1 Check if the sensor is damaged or not connected properly to the device 2 Ch...

Page 21: ...sories must be disposed of in accordance with the applicable national provisions for electronic components In accordance with the requirements of EU Directive 2012 65 EC Waste Electrical and Electroni...

Page 22: ...to OD 7 8 2 of reading 80 ml min up to OD 1 1 8 2 of reading 200 ml min up to OD 1 5 8 2 of reading 400 ml min The specified accuracies were determined under the following conditions Medium Water Med...

Page 23: ...9 16 13433 PCT 3 8 x 1 8 0 017 2 5 0 100 10 3 8 1 8 5 8 13422 PCT 1 2 x 3 32 0 017 2 5 0 100 16 9 2 40 1 2 3 32 11 16 13428 PCT 1 2 x 1 8 0 017 2 5 0 100 16 9 2 40 1 2 1 8 3 4 13430 PCT 1 2 x 3 16 0 0...

Page 24: ...1 2 x 3 16 PCT 3 4 x 1 8 PCT 3 4 x 3 16 PCT 1 x 1 8 PCT 1 x 3 16 PCT 1 1 4 x 3 16 Size H x W x D 85 x 80 x 10 6 mm 100 x 80 x 10 6 mm Weight 28 g 31 g Material polyamide polyamide Mounting four M5 scr...

Page 25: ...nue of approximately 7 billion We deliver innovative equipment and components consumable supplies aftermarket parts software and digital solutions and support services through five operating segments...

Reviews: