5.4. Combined meter

The combined meter measures the fill water quality and the throughput (l/

min and total throughput).

 

Mains plug 

indication

 

 

Total 

 

treated water

 

 

Current 

throughput

 

 

Water 

 

temperature

 

 

Switch between 

litres/gallons

 

 

ON/OFF 

 

automatic mode

 

Automatic 

 

quality 

 

monitoring

 

 

Battery

 

 indication

 

 

Limit

 

 

Current 

conductivity value

 

 

Set the limit

 

 

Switch between 

µS TDS

Set key

Each time the set key is pressed, the limit is increased by 10 µS/cm or 6 TDS. 

Hold the key for 3 seconds to reset the limit to zero. Programming the limit 

ensures that a warning is issued when the ion exchange resin is spent. If 

the set limit is exceeded, the backlight flashes red and an acoustic alarm 

sounds. In addition, the floating contact is switched over (only when using the 

optional power supply unit).

Unit key

Use this key at any time to switch between electrical conductivity (µS/cm) 

and TDS (total dissolved solids) as well as between °C and °F. The first two 

are units of measurement for the amount of minerals dissolved in the water. 

Most European component manufacturers use the unit of measurement 

µS/cm (microsiemens per centimetre). Water hardness at the output of 

the refill device can be determined using the following rule of thumb: 1 °fH 

Summary of Contents for PUROTAP easy II

Page 1: ...www elysator com Installation Function Operation Maintenance Operating instructions PUROTAP easy II...

Page 2: ......

Page 3: ...description 6 4 2 Dimensions 6 4 3 Performance data 6 5 Operation 5 1 1 Draining spent resin 7 5 1 2 Filling the resin canister 7 5 2 Connection 8 5 3 Commissioning 8 5 4 Combined meter 9 6 Maintenanc...

Page 4: ...4 1 Safety instructions Not drinking water Maximum 60 C Maximum pressure 6 bar not constantly Flow rate max 15 l min 6 bar 15 l min...

Page 5: ...e resin The device has an integral meter that constantly measures and displays fill volume throughput and water quality There is no need for an external power supply and no chemicals are released into...

Page 6: ...d mesh filter 5 Head with immersion pipe 6 PUROTAP LFM 20 combined meter 7 Handle 8 Fine mesh filter 9 Outlet 4 2 Dimensions Dimensions PUROTAP easy II Height 74 0 cm Width 32 5 cm Depth 25 0 cm Weigh...

Page 7: ...n drain valve 4 Open shut off valve 2 5 Flush until no more resin is left in the canister 6 Remove collection bag from the drain and allow to drip 7 Close drain valve and screw round mesh filter with...

Page 8: ...pipe is screwed on tightly 5 2 Connection Connect the device to the raw water at the inlet 1 using an adequately sized hose Connect the outlet 9 to the system to be filled also using a hose The device...

Page 9: ...ogramming the limit ensures that a warning is issued when the ion exchange resin is spent If the set limit is exceeded the backlight flashes red and an acoustic alarm sounds In addition the floating c...

Page 10: ...ptional power supply unit the combined meter does not automatically switch off and cannot be switched off manually If the combined meter has switched off automatically it will automatically restart on...

Page 11: ...e 4 screws on the red cover lift off the cover and insert 3 new AAA LR03 1 5 V batteries Make sure the cover seal is seated properly Refit the cover and secure with the screws Check the device is work...

Page 12: ...nt 7 Spare parts Item no Designation 100 967 PUROTAP easy II complete immersion pipe 100 620 PUROTAP easy II complete head 102 278 PUROTAP LFM 20 combined meter 100 255 PUROTAP easy II O ring for head...

Reviews: