corresponds to approx. 20 µS/cm or 1 °dH corresponds to approx. 35 µS/cm. 

 

Sequence of units: µS/cm -- °C | µS/cm -- °F | TDS -- °C | TDS -- °F

ON/OFF key (MANU/AUTO)

Press the   key once to switch on the combined meter. The water quality is 

measured for 10 seconds and compared to the set limit. During measuring, 

the screen backlighting is blue and the conductivity value is displayed. If the 

measured value is below the set limit, everything is OK. If the measured value 

exceeds the limit, the display flashes red, an alarm signal sounds and the 

floating contact switches over (only when using the optional power supply 

unit). The current throughput and the water temperature, as well as the 

total water throughput, are continuously displayed. Press the key for approx. 

3 seconds to switch off the device. If the combined meter is switched on and 

measures no throughput for about one hour, it automatically switches off. 

During operation with the optional power supply unit, the combined meter 

does not automatically switch off and cannot be switched off manually. If the 

combined meter has switched off automatically, it will automatically restart 

once the throughput (> 2 l/min) resumes. If the combined meter is switched 

off using the ON/OFF key, it will not restart automatically. 

Auto mode

Press the   key

 

twice in quick succession to start automatic monitoring by 

the combined meter. The 

 symbol is shown to indicate that monitoring 

has been enabled. In auto mode, the meter only takes measurements when 

water is actually passing over it. If the throughput is interrupted, the meter 

continues to show the last captured value. When a throughput is present, the 

meter takes a new reading of the conductivity value after every 10 litres and 

displays it. If the limit is exceeded in two successive measurements, the 

display continuously flashes red and an acoustic alarm sounds. In addition, 

the floating contact is switched over. If the   key is pressed for a third time, 

the meter exits auto mode. Auto mode is recommended for operation with 

an automatic topping up device.

Summary of Contents for PUROTAP easy II

Page 1: ...www elysator com Installation Function Operation Maintenance Operating instructions PUROTAP easy II...

Page 2: ......

Page 3: ...description 6 4 2 Dimensions 6 4 3 Performance data 6 5 Operation 5 1 1 Draining spent resin 7 5 1 2 Filling the resin canister 7 5 2 Connection 8 5 3 Commissioning 8 5 4 Combined meter 9 6 Maintenanc...

Page 4: ...4 1 Safety instructions Not drinking water Maximum 60 C Maximum pressure 6 bar not constantly Flow rate max 15 l min 6 bar 15 l min...

Page 5: ...e resin The device has an integral meter that constantly measures and displays fill volume throughput and water quality There is no need for an external power supply and no chemicals are released into...

Page 6: ...d mesh filter 5 Head with immersion pipe 6 PUROTAP LFM 20 combined meter 7 Handle 8 Fine mesh filter 9 Outlet 4 2 Dimensions Dimensions PUROTAP easy II Height 74 0 cm Width 32 5 cm Depth 25 0 cm Weigh...

Page 7: ...n drain valve 4 Open shut off valve 2 5 Flush until no more resin is left in the canister 6 Remove collection bag from the drain and allow to drip 7 Close drain valve and screw round mesh filter with...

Page 8: ...pipe is screwed on tightly 5 2 Connection Connect the device to the raw water at the inlet 1 using an adequately sized hose Connect the outlet 9 to the system to be filled also using a hose The device...

Page 9: ...ogramming the limit ensures that a warning is issued when the ion exchange resin is spent If the set limit is exceeded the backlight flashes red and an acoustic alarm sounds In addition the floating c...

Page 10: ...ptional power supply unit the combined meter does not automatically switch off and cannot be switched off manually If the combined meter has switched off automatically it will automatically restart on...

Page 11: ...e 4 screws on the red cover lift off the cover and insert 3 new AAA LR03 1 5 V batteries Make sure the cover seal is seated properly Refit the cover and secure with the screws Check the device is work...

Page 12: ...nt 7 Spare parts Item no Designation 100 967 PUROTAP easy II complete immersion pipe 100 620 PUROTAP easy II complete head 102 278 PUROTAP LFM 20 combined meter 100 255 PUROTAP easy II O ring for head...

Reviews: