The loader electronic board contains a buzzer for signalizing errors. The number of beeps in a
row specifies the problem according to the table below.
=== Ende der List e f ür T ext mar ke I nhalt ===
No. of
beeps
Error description
1
The beam of the vertical laser cannot be detected.
Defective laser or detector, dirt in between, arm in between.
2
The beam of the vertical laser is continuously detected.
Error of the fing
er’s motor.
3
No crucible on the loader although the loader received the command to load the next crucible.
4
The beam of the horizontal laser cannot be detected.
Defective laser or detector, crucibles don’t move due to blockage, defective motor, etc.
5
The furnace is not opened for the loader to place a crucible on to the pedestal.
Error in the piston control or piston sensor.
6
The fingers didn’t take the crucible. The crucible blocks in its tray, no compressed air, faulty laser or
sensor, etc.
7
The loader received the command for loading but the furnace is not open, or the piston came down with
big delay.
8
One dispenser empty.
9
One dispenser empty.
∞
Infinitive number of beeps in manual mode.
Piston sensor disadjusted (low end of furnace cylinder).