Checklist for Insulation Tests
Page 13
FAULT
Buzzer sounds continuously when the test button is pressed.
CAUSE
REMEDY
No or incorrect earthing connection
(valid only for testers with plug-in earth connection:
ISOTEST Type II T / II RT from Serial No. 14796
ISOTEST Type VT / C-VRT from Serial No. 21 000
or from October 1991
The earthing plug must be pushed
into the earthing socket up to the
stop. There must be no material
from the leather carrying case
between plug and socket.
Idle - no electrode contact with the test
specimen
Battery charge too low
ELMED Isotest testers are equipped with
exhaustive discharge protection
Replace battery or inspect old
battery and recharge.
During the test - with test electrode
in contact with the test specimen
Excessive energy flow e.g. due to
moisture
Reduce the display sensitivity at
the adjustment knob until the
continuous signal stops. The
reduction in the display sensitivity
has no effect on the level of the set
voltage!!
In other cases, a "continuous buzzer" is the signal for a detected flaw in the
test specimen!