background image

 
ENGLISH 

 

Putting into Operations

 

© Elma GmbH & Co KG 

BA/Super-Elite/GB/0104 

5

4 Putting 

into 

Operations 

1.  Compare voltage on type-plate (1) at the machine-hood (2) 

with the voltage in your workshop. 

Preparation

2.  Place the unit onto a stable and dry surface 

3.  Set up the Machine with the heating tube (3) facing toward 

you and unscrew the transport safety (5) in the bracket (6). 

4.  Turn the black turning handle (7) to the front, move the 

bracket (6) upward till stop is reached. – By turning the 
handle (7) to the back the bracket (6) is fixed. 

5.  Fill the glass jars (8) in the following order:  

Fill glass jars (8) to the left of the heating tube (3) with elma 
cleaning concentrate 1:9 till the first level mark. Dilute up to 
the end of the hammered mark with ordinary tap-water. 
Fill the second glass jar with distilled water and the 
remaining third one with elma suprol spezial. 

Filling the jars

After 50-60 cleanings the solution should be renewed at all 
events.  
The rinsing-solution in the last glass jar must always be kept 
clean. 

Working with the Elma Super Elite 

5.1 

Placing the watch parts to the cleaned 

1.  Put your right hand round the cleaning basket (10) whilst 

detaching with your left hand by a light turn to the left, the 
basket flange (11) with suction blades (11) form the 
cleaning basket. 

2.  Push out the 3 basket insets in the body with your right 

forefinger into the hollow of your left hand. Then place the 
watch parts – which must be kept separate according to 
size and degree of sensitiveness – into the various insets. 
Place these inset again in a body into the basket with the 
help of your left forefinger.  
The small round box with cover in the middle of the inset is 
for the smallest parts. 

3.  Attach the basket by slightly turning the flange to the right, 

and lock the basket with the lock decide provided on the 
flange plate. 

4.  Plug in the mains cable to a earth protective mains socket. 

 

 

Summary of Contents for Super Elite

Page 1: ...3 jars with lid Adjustable rotational speed screen basket Timer 1 basket frame Heating for drying chamber seperatly switchable 1 basket lid 2 insert 5 div h 15mm 1 insert 1 div h 15mm 1 Miniature bask...

Page 2: ...ENGLISH Operating Instructions Super Elite Watch Cleaning Machine...

Page 3: ...al 7 8 Manufacturer s contact address 8 1 General The present Operating Instructions are part of the delivered equipment They must be ready for use at any time and remain with the unit in case of resa...

Page 4: ...uch the cover at the plastic handle only The bottom of the drying chamber can reach temperatures exceeding 90 C depending on the period of operation Hot surface In order to prevent electrical accident...

Page 5: ...l details Mains voltage as per version Vac 100 120 or 240 240 Mains frequency Hz 50 60 Power consumption of motor Watt 18 Power consumption of heating Watt 200 Outer dimensions of unit D H mm 330 500...

Page 6: ...suprol spezial Filling the jars After 50 60 cleanings the solution should be renewed at all events The rinsing solution in the last glass jar must always be kept clean 5 Working with the Elma Super El...

Page 7: ...s about 3 to 5 minutes depending on the extent to which the parts are soiled 3 Start the motor 13 by means of the switch 14 separately provided 4 Set the speed regulator to slow with the aid of the bl...

Page 8: ...own safety we recommend that you check the mains cable and the mains connection jack at the unit for damage at regular intervals Mains cable and mains connection In case of technical faults please co...