background image

 

Elma 1350B 

side 10 

 
 

Name and Functions 

 
1. Windscreen 
If you operate at wind speed over 
10m/sec, please put protective 
accessories in front of the 
microphone. 
 
2. Display 
 
 
 
 
 
 
 
 
 
 
 
 
 
 3. Power Button 
The key (3) turns the sound level meter ON or OFF. 
 
4. MAX / MIN Hold button 
Press           button to enter the maximum and minimum   recording mode. Select the 
proper Level range before using MAX/MIN to ensure that reading value will not exceed the 
measurement range. Press once to select MAX value. Press again to select MIN value, 
and press again to select current Value with “ MAX MIN ”  annunciator blinking.   
Press           and hold down button for 2 seconds to exit the MAX/MIN mode. 
 
Note: If change sound level range or change A-C weight, the MAX.MIN mode will 
clear.  
 
5. Frequency Weighting Select Button 
A: A - Weighting. for general sound level measurements. 
C:

 

C - Weighting. for checking the low-frequency content of noise. 

( If the C-Weighted level is much higher than the A-weighted level, then there is a large 
amount of low-frequency noise ) 
 
6. Time Weighting Select Button. 
FAST : for normal measurements 
SLOW : for checking average level of fluctuating noise. 
 

MAX

A / C

LEVEL

FAST

SLOW

130 dB

80 dB

100 dB

Lo  =

Hi   =

Med=

32

80

50

IEC  651  TYPE  II

MIN

C

A

L

O

U

T

P

U

T

OPEN

TERY

NEDA 1604  6F22  006P

9V  BAT

4

7

5

6

3

2

10

11

9

8

1

Low battery indication

Minmum indication

Maximum indication

Over range

Fast response

Sound Level readout

Range Indicate

C-Weighting

A-Weighting

Slow response

Under range

MAX

MIN

MAX

MIN

Summary of Contents for 1350B

Page 1: ...Elma 1350B Dansk norsk vejledning Side 3 5 Svensk bruksanvisning Sida 5 7 English usermanual Page 8 12 DK 63 98 157 229 SE 42 104 19 NO 80 227 66 EAN 5703317410283...

Page 2: ...nformation 5 Specifikationer 5 Beskrivning 6 Display 6 Anv ndning 6 Kalibrering 7 Byte av batteri 7 Egna anteckningar 7 English usermanual 8 Safety Information 8 Environment Conditions 8 Maintenance C...

Page 3: ...0 1dB N jagtighed 1 5dB ref 94dB 1kHz Mikrofon Elektrisk kondensator mikrofon Arbejdes omgivelser 0 40C 10 90 RH Lager 10 t il 60 C 10 75 RH Standarder IEC 651 type 2 og ANSI S1 4 type 2 Udgang 1V rm...

Page 4: ...SLOW for gennemsnitsm ling p st rkt varierende niveauer V lg A eller C v gtning af m lingen 5 A for normal m ling og C for lavfrekvens m ling af f eks maskinst j Hvis dBC v rdien er st rre end dBA v r...

Page 5: ...et nyt 9V batteri Svensk manual S kerhetsinformation L s denna manual innan anv ndning av instrumentet Kom inte i direkt kontakt med elektrisk sp nning Undvik smutsiga milj er Denna manual kan ndras...

Page 6: ...g batterisp nning Avl sning av m tv rde M tning i dBA M tning i dBC UNDER indikerar att m tt v rde r under det valda m tomr det M tning r VER valt omr Visning av valt m tomr de Anv ndning Sl p instru...

Page 7: ...kopplas funktionen ur och displayen visar gonblicksv rden igen Efter val av funktioner kan den aktuella ljudniv n avl sas p displayen Instrumentet sl s av genom att trycka p den r da knappen 3 igen K...

Page 8: ...th EMC When servicing use only specified replacement parts General Description Thank you for using our Sound Level Meter To ensure that you can get the most from it we recommend that you read and foll...

Page 9: ...limit of range UNDER is when input is Less than lower limit of range MAX MIN Hold Hold readings the Maximum and Minimum Value AC Output 1 Vrms at FS full scale Output impedance Approx 100 Ohms FS mean...

Page 10: ...e MAX MIN mode Note If change sound level range or change A C weight the MAX MIN mode will clear 5 Frequency Weighting Select Button A A Weighting for general sound level measurements C C Weighting fo...

Page 11: ...ings Display dBA Time Weighting FAST Measurement Mode MAX MIN Mode function disable Level Range 50 to 100dB 2 Insert the microphone housing carefully into the insertion hole of the calibrator 3 Turn o...

Page 12: ...install a 9V battery in the battery compartment 2 Turn power ON and select the desired response time and weighting If the sound source consists of short bursts or only catching sound peak set respons...

Reviews: