background image

 

 

Kewprove 3 

 4 

 

English user manual 
Kewprove 3 proving unit 

Warning 

Caution, risk of electric shock. Instructions must be followed to avoid danger to 
persons. 

This unit puts out a hazardous voltage and must only be used by a qualified person and in 
accordance with these instructions. If the equipment is not used in the manner described 
the protection provided by the equipment may be impaired. 
Check that the instrument is free from moisture, grease and dust before use. 
Do not use if the unit is damaged in any way. 
Do not use in wet conditions. 
Clean only with a dry cloth do not use solvents. 
There are no user replaceable parts inside 

– do not open other than for replacing batteries. 

To access the batteries, undo the two screws at the rear of the unit and remove cover. 

 

Batteries should be disposed of with compliance to current recycling and disposal 
regulations. 

 

To ensure the safety of personnel working on electrical circuits, BEST PRACTICE 
calls for the use of a proving unit before working on dead circuits. 

  50 

– 690V proving unit with just two terminals to test AC and DC voltage detectors 

  Fast response and works with low impedance voltage testers 

  Ramps down through voltage levels 

  LED indication of functioning unit 

Instructions 

1.  Test your voltage detector on the proving unit, by placing the 

probes into the front terminals. 

2.  The LED on the proving unit should light up and your voltage 

detector should indicate voltage present. 

3.  Continue to hold your voltage detector in the front terminals of 

the Kewprove 3 and the voltage level output will fall in steps to 
400V, 230V, 100V & 50V to prove these levels of voltage. 

4.  Prove your voltage detector is still working by placing the probes 

into the front terminals again. 

5.  The LED on the proving unit should light up and your voltage 

detector should indicate voltage present. 

6.  Continue to hold your voltage detector in the front terminals of 

the Kewprove 3 and the voltage level output will fall in steps to 
400V, 230V, 100V & 50V to prove these levels of voltage. 

Specifications 

Output voltage:  

 

180V in to 6.8K

Ω, 720V - 820V in to 10MΩ 

Operating conditions: 

 

-10 to 40°C, 70% RH 

Batteries:    

 

6 x AA Alkaline 

 
 
 
 
 
 