PC computer interface capabilities 

The instrument equips with an optical isolated interface port at the meter back for data communication. 
Press 

HOLD 

button while turning the meter on to enable meter RS232 output. Optional purchase PC 

interface kit BRUA20X is required to connect the meter to the PC computer RS232 or USB ports. 

MAX/MIN at Fast 20/s measurement mode 

Press REC button momentarily to activat

e MAX/MIN recording mode. The LCD “MAX MIN” turn on, and 

the reading update rate will be increased to 20/second. The meter beeps when new MAX (maximum) or 
MIN (minimum) reading is updated.  Press the button momentarily to read the MAX and MIN readings in 
sequence. Press the button for 1 second or more to exit MAX/MIN recording mode. Auto-ranging 
remains, and Auto-Power-Off is disabled automatically in this mode. 

5ms CREST capture mode  

Press CREST button momentarily to activate CREST (Instantaneous Peak-Hold) mode to capture 
voltage or current signal duration as short as 5ms. The LCD “C” & “MAX” turn on. The meter beeps when 
new MAX (maximum) or MIN (minimum) reading is updated. Press the button momentarily to read the 
MAX and MIN readings in sequence. Press the button for 1 second or more to exit CREST capture mode. 
Auto-ranging and Auto-Power-Off are disabled automatically in this mode. 

Backlighted LCD display 

Press the SELECT button for 1 second or more to toggle the LCD backlight. The backlight will also be 
turned off automatically after 32 seconds to extend battery life. 

Hold 

The hold feature freezes the display for later view. Press the HOLD button momentarily to toggle the hold 
feature. 

Relative Zero mode 

Relative zero allows the user to offset the meter consecutive measurements with the displaying reading 
as the reference value. Practically all displaying readings can be set as relative reference value including 
MAX/MIN feature readings. Press the REL button momentarily to toggle relative zero mode. 

Manual or Auto-ranging 

Press the RANGE button momentarily to select manual-ranging, and the meter will remain in the range it 
was in, the LCD 

 turns off.  Press the button momentarily again to step through the ranges.  Press 

and hold the button for 1 second or more to resume auto-ranging. 

Note: Manual ranging feature is not available in Hz and 

 functions. 

Set Beeper Off 

Press the RANGE button while turning the meter on to temporarily disable the Beeper feature. Turn the 
rotary switch OFF and then back on to resume. 

Beep-

Jack™ Input Warning 

The meter beeps as well as displays “InEr” to warn the user against possible damage to the meter due to 
improper connections to the 

A, mA, or A input jacks when other function (like voltage function) is 

selected. 

Auto-Power-Off (APO) 

The Auto-Power-Off (APO) mode turns the meter off automatically to extend battery life after 
approximately 34 minutes of no rotary switch or push button operations. To wake up the meter from APO, 
press the SELECT, CREST or REC button momentarily or turn the rotary switch OFF and then back on. 
Always turn the rotary

 switch to the OFF position 

when the meter is not in use 

Summary of Contents for 5706445410286

Page 1: ...MANUAL Elma BM251s 257s Dansk Norsk 5 13 Svensk 14 22 English 23 34 EAN 5706445410286 Elma BM251s 5706445410293 Elma BM257s...

Page 2: ...rm 8 Sp gelsessp ndinger 8 DCV ACV linjefrekvens funktioner 8 modstand gennemgang kapacitans diode test funktioner 9 DCmV ACmV temperatur C F funktioner 9 A mA og A funktioner 10 Elektrisk felt detekt...

Page 3: ...larm 17 Ghostvoltage 17 DCV ACV linjefrekvensfunktioner 17 resistans genomg ng kapacitans diodtestfunktion 18 DCmV ACmV temperatur C F funktioner 18 A mA och A funktioner 19 Elektriskt f ltdetektering...

Page 4: ...ACV Line Frequency functions 26 Resistance Continuity Capacitance Diode test functions 27 DCmV ACmV Temperature oC oF functions 27 A mA and A Current functions 28 Electric Field EF Detection 28 PC com...

Page 5: ...el brug og andet udstyr heri f eks station re motorer med permanent forsyning til fast installation M lekategori II Kat II Denne kategori er for m linger udf rt p ledningssystemer direkte forsynet til...

Page 6: ...under justeringer Gennemsnits RMS kalibrering RMS Root Mean Square er en norm som bruges til at beskrive den effektive eller tilsvarende DC v rdi p et AC signal De fleste digitale multimetre er med R...

Page 7: ...ner p 60dB ved DC til 60Hz i ACV funktion og 120dB ved DC 50 og 60Hz i DCV funktion Hvis hverken NMRR eller CMRR specifikationerne er udpenslet vil et digitalt multimeters udf rsel v re un jagtig Funk...

Page 8: ...Autocheck funktionen p str mkredse som kan blive beskadiget ved s lave impedanser I stedet bruges omskifterfunktionen eller impedanssp ndings funktionen for at minimere ladning for s danne kredsl b DC...

Page 9: ...ugbar for kontrol af ledningsforbindelser og funktion af afbrydere En kontinuerligt bip tone indikerer en komplet ledning system Et normalt forudindtaget sp ndingsfald for en god diode er mellem 0 4V...

Page 10: ...arbejdssp ndingen for beskyttelsessikringerne Elektrisk felt detektering Ved volt eller str m funktionen tryk da p EF knappen i et sekund eller mere og slip for at bladre imellem EF beskyttelses funk...

Page 11: ...m lingerne Tryk p knappen i et sekund eller mere for at returnere fra denne funktion Auto omr de funktionen forbliver og Auto sluk funktionen bliver deaktiveret automatisk i denne funktion Baggrunds...

Page 12: ...l bet fjern testledningerne og sluk instrumentet f r der foretages vedligeholdelse p instrumentet Brug aldrig instrumentet n r dette er bent Is t altid kun samme type sikring Reng ring og opbevaring P...

Page 13: ...EN61000 4 4 EN61000 4 5 EN61000 4 6 EN61000 4 8 EN61000 4 11 I et RF felt p 3V m kapacitans funktion er ikke specificeret Andre funktionsomr der Total n jagtighed specificeret n jagtighed 100 digits...

Page 14: ...versp nningsskategori III apparatur som r monterad i den fasta installationen Notera Ex Matare brytare korta grenkontakter ljus system apparatur monterat i industriinstallationer versp nningsskategor...

Page 15: ...nomsnittlig RMS kalibrering RMS Root Mean Square r en term som anv nds f r att beskriva det effektiva eller ekvivalenta v rdet av en AC signal De flesta digitala multimetrar anv nder average sensing R...

Page 16: ...ns m jlighet att ta bort sp nningar som kan f rorsaka att displayen pendlar eller ger en felvisning Denna serie har CMRR specifikationer p h gre n 60dB fr n DC till 60Hz i ACV funktionen 120dB vid DC...

Page 17: ...1000VAC kan vara upp till 566mA 1000V x 1 414 2 5k f r att pl tsligt reduceras till ca 3 8mA 1000V x 1 414 375k vid en br kdel av en sekund Anv nd aldrig Autocheck funktionen p str mkretsar som kan t...

Page 18: ...nv nds f r kontroll av ledningsf rbindelser och funktion av str mbrytare En kontinuerligt pip ton indikerar en komplett ledning system Ett normalt sp nningsfall f r en bra diod r mellan 0 4V till 0 9V...

Page 19: ...unktionen tryck d p EF knappen i en sekund eller mera f r att v xla till EF detekteringsfunktionen Instrumentet visar nu E F Signalstyrkan visas nu som ett stapeldiagram p displayen samt att man f r v...

Page 20: ...ra f r att t nda bakgrundsbelysningen Den sl cks automatiskt efter 32 sekunder f r att spara p batterierna Hold funktion Hold funktionen fryser det sista uppm tta v rdet Tryck p HOLD knappen momentvis...

Page 21: ...entet med j mna mellanrum med en fuktad trasa med lite milt reng ringsmedel absolut inga slipande eller uppl sande medel Om instrumentet inte skall anv ndas under en l ngre period ca 60 dagar tag d ur...

Page 22: ...61000 4 3 EN61000 4 4 EN61000 4 5 EN61000 4 6 EN61000 4 8 EN61000 4 11 I RF omr det p 3V m anges inte kapacitansfunktionen Andra funktionella omr den Total noggrannhet specificerad noggrannhet 100 dig...

Page 23: ...ndustrial use and some other equipment for example stationary motors with permanent connection to the fixed installation Measurement Category II CAT II is for measurements performed on circuits direct...

Page 24: ...ty contacts identifying potentiometer clicks and indicating signal spikes during adjustments Average sensing RMS calibrated RMS Root Mean Square is the term used to describe the effective or equivalen...

Page 25: ...effect that can cause digit rolling or offset in voltage measurements This series has a CMRR specifications of 60dB at DC to 60Hz in ACV function and 120dB at DC 50 and 60Hz in DCV function If neithe...

Page 26: ...C for example can be up to 566mA 1000V x 1 414 2 5k decreasing abruptly to approx 3 8mA 1000V x 1 414 375k within a fraction of a second Do not use AutoCheckTM mode on circuits that could be damaged b...

Page 27: ...eep tone indicates a complete wire Normal forward voltage drop forward biased for a good silicon diode is between 0 400V to 0 900V A reading higher than that indicates a leaky diode defective A zero r...

Page 28: ...Field EF Detection At Volt or Current function press the EF button for one second or more and release to toggle to EF Detection feature The meter displays E F when it is ready Signal strength is indic...

Page 29: ...urned off automatically after 32 seconds to extend battery life Hold The hold feature freezes the display for later view Press the HOLD button momentarily to toggle the hold feature Relative Zero mode...

Page 30: ...ck operating procedure as described in this user s manual If the instrument voltage resistance input terminal has subjected to high voltage transient caused by lightning or switching surge to the syst...

Page 31: ...0 3 2 EN61000 3 3 EN61000 4 2 EN61000 4 3 EN61000 4 4 EN61000 4 5 EN61000 4 6 EN61000 4 8 EN61000 4 11 In an RF field of 3V m Capacitance function is not specified Other function ranges Total Accuracy...

Page 32: ...Input Impedance Initially approx 2 5k 120pF nominal Impedance increases abruptly within a fraction of a second as display voltage is above 50V typical Ended up impedances vs display voltages typicall...

Page 33: ...000V 1 0 3d Test Current 0 56mA typically Open Circuit Voltage 1 8VDC typically DC Current RANGE Accuracy Burden Voltage 600 0 A 0 5 5d 0 10mV A 6000 A 0 5 3d 60 00mA 0 5 5d 1 7mV mA 600 0mA 0 5 3d 6...

Page 34: ...0Hz 10kHz 6A 1A 10Hz 1kHz 10A 6A 10Hz 1kHz Accuracy 0 03 3d Non Contact EF Detection Typical Voltage Bar Graph Indication 20V tolerance 10V 36V 55V tolerance 23V 83V 110V tolerance 59V 165V 220V toler...

Page 35: ......

Page 36: ...20131017MD v1 0 20220708 hlo v 1 1...

Reviews: