background image

Page 7

98557C (Rev. J - 8/14)

EZWSMD*1A     EZWSM8*1A     EZWSM8*2A     LZWSMD*1A     LZWSM8*1A     LZWSM8*2A

VERIFY CONTROL BOARD  SOFTWARE

1) To verify the software program of the control board the
    unit will need to be shut down and restarted. The chiller
    (if present) does not need to be shut down and restarted.
2) The units lower panel must be open to access the power
    cord and wall outlet.
3) Shut down the unit by unplugging the power cord from the
    wall outlet.
4)  Restart the unit by plugging the power cord back into the
    wall outlet.
5) Upon start up the bottle count display will show the
    software designation of BF6, BF7, BF8, BF9 or BF11.
6) Reference the BF6-BF7-BF8-BF9 or BF11 instructions for
    setting the control board.

ACCESSING THE PROGRAMMING BUTTON

1)   To access the program button the lower panel of the
      unit must be must be opened. The programming button is
      located at the bottom right corner of the upper panel.
     This area of the unit is concealed by the lower panel.

RESET THE FILTER MONITOR

1) Instructions apply to filtered units only.
2) Depress the program button for approximately 2 seconds
    until the display changes then release. The display will
    change and scroll through two messages:
      

 “RST FLTR” – Reset Filter Monitor

     “SETTINGS” – System Settings Sub Menu
    If the program button is not pushed again the display
    will scroll through the two messages above for
    three cycles and then default back to bottle count and
     be back in run mode.
3) When the display changes to “RST FLTR”, depress
    the button again. The display will change to show
     “FLTR =”. Depress the button again and the display
    will show “FLTR =0”
4) The Green LED should be illuminated indicating that

    the visual filter monitor has been reset.

 SETTING RANGE OF THE IR SENSOR

1) Depress the program button for approximately 2 seconds
    until the display changes then release. The display will
    change and scroll through two messages:
       “RST FLTR” – Reset Filter Status LED
     “SETTINGS” – System Settings Sub Menu
    If the program button is not pushed again the display
    will scroll through the two messages above for
    three cycles and then default back to bottle count
     and be back in run mode.
2) When the display changes to “SETTINGS”, depress
    the button again. The display will change to show
     “RNG SET“-  Range set for IR sensor.
      “UNIT TYP“ - Type of unit (REFRIG or NON-RFRG)
     “RST BCNT“ - Reset bottle count
3)  When display shows “RNG SET” push program
    button once the display will show current value
     (can be 1 – 10) e.g. “RNG = 3”.
4) Once display shows current value push the
    program button to scroll through value of 1 – 10.
    Select the desired range setting.
5) Once range is selected allow approximately
    4 seconds to pass and then the display will go
     back to bottle counter and be in run mode.
6)  Test bottle filler by placing bottle or hand in front

    of sensor to make sure water is dispensed.

SETTING UNIT TYPE

1) Depress the program button for approximately 2 seconds
    until the display changes then release. The display will
    change and scroll through two messages:
      “RST FLTR” – Reset Filter Status LED
     “SETTINGS” – System Settings Sub Menu
    If the program button is not pushed again the display
    will scroll through the two messages above for
    three cycles and then default back to bottle count
      and be back in run mode.
2)  When the display changes to “SETTINGS”, depress
    the button again. The display will change to show
     “RNG SET“-  Range set for IR sensor.
       “UNIT TYP“ - Type of unit (REFRIG or NON-RFRG)
      “RST BCNT“ - Reset bottle count
3)  When display shows “UNIT TYPE” push program
    button once the display will show current value
       Can be REFRIG or NON-RFRG
4) Push button once to change value. Once value is
      selected the display will show the new value.
       (Can be REFRIG or NON-RFRG)
     “REFRIG“ - stands for refrigerated product. In this
     setting the flow rate is estimated at 1.0 gallon per minute.
     “NON-RFRG“ - stands for nonrefrigerated product.
    In this setting the  flow rate is estimated
      at 1.5 gallons per minute.
      Both “REFRIG“ and “NON-RFRG“ simulate
       1 bottle equal to 20 oz.
5)  Allow approximately 4 seconds to pass and the display
     will return to bottle counter and be in run mode.

    

 RESETTING BOTTLE COUNT

1) Depress the program button for approximately 2 seconds
    until the display changes then release. The display will
    change and scroll through two messages:
      “RST FLTR” – Reset Filter Status LED
      “SETTINGS” – System Settings Sub Menu
    If the program button is not pushed again the display
    will scroll through the two messages above for
    three cycles and then default back to bottle count
     and be back in run mode.
2)  When the display changes to “SETTINGS”, depress
    the button again. The display will change to show
     “RNG SET“-  Range set for IR sensor.
       “UNIT TYP“ - Type of unit (REFRIG or NON-RFRG)
      “RST BCNT“ - Reset bottle count
    If the button is not pushed again the display will scroll
    through the three messages above for the cycles and
     return to run mode.
3)  When display shows “RST BCNT” push program button
    once the display will show current value e.g. “00033183”.
4) Once display shows current value push the program
    button once more to reset back to 0. The display will
    show BTLCT = 0 for approximately 2 seconds and
    then return to run mode showing 00000000 bottles.
5)  Testing the bottle counter:
     REFRIG units: Place bottle or hand in front of sensor
       for 9.4 seconds to see bottle counter count 00000001.
         (This is based on filling a 20 oz. bottle)
     NON-RFRG units: Place bottle or hand in front of sensor
       for 6.25 seconds to see bottle counter count 00000001.
         (This is based on filling a 20 oz bottle)

BF9 PROGRAM

SETTING THE CONTROL BOARD

Summary of Contents for LZWSM8K

Page 1: ...WATER LINES IS A COMMON PROCEDURE THIS GROUNDING MAY BE IN THE BUILDING OR MAY OCCURAWAY FROM THE BUILDING THIS GROUNDING CAN CAUSE ELECTRICAL FEEDBACK INTOAFOUNTAIN CREATINGAN ELECTROLYSIS WHICH CAUS...

Page 2: ...98557C Rev J 8 14 Page 2 EZWSMD 1A EZWSM8 1A EZWSM8 2A LZWSMD 1A LZWSM8 1A LZWSM8 2A FIG 3 LZWSM8 LZWSMD EZWSM8 EZWSMD ROUGH IN...

Page 3: ...Page 3 98557C Rev J 8 14 EZWSMD 1A EZWSM8 1A EZWSM8 2A LZWSMD 1A LZWSM8 1A LZWSM8 2A LZWSM8P LZWSMDP EZWSM8P EZWSMDP ROUGH IN FIG 4...

Page 4: ...er supply valve and union not provided Turn on the water supply and flush the line thoroughly 4b Make water supply connections for Non Fltered units Install a shut off valve and union connection to bu...

Page 5: ...Bottle Filler the LCD Bottle Counter should illuminate 15 Verify proper dispensing by placing cup hand or any opaque object in front of sensor area and verify water dispenses Note the first initial d...

Page 6: ...green LED should now be illuminated indicating that the visual filter monitor has been reset SETTING RANGE OF THE IR SENSOR 1 Depress the program button for approximately 2 seconds until the display c...

Page 7: ...play will go back to bottle counter and be in run mode 6 Test bottle filler by placing bottle or hand in front of sensor to make sure water is dispensed SETTING UNIT TYPE 1 Depress the program button...

Page 8: ...ed allow approximately 4 seconds to pass and then the display will go back to bottle counter and be in run mode 6 Test bottle filler by placing bottle or hand in front of sensor to make sure water is...

Page 9: ...fault back to bottle count and be back in run mode 2 When the display changes to SETTINGS depress the button again The display will change to show RNG SET Range set for IR sensor UNIT TYP Type of unit...

Page 10: ...ea Fig 8 LISTE DES PI CES DU FILTRE Voir Fig 8 1 2 3 4 5 6 7 WATER FILTER EXPLODED VIEW FIG 8 PRINTED IN U S A IMPRESO EN LOS E E U U IMPRIM AUX U FORPARTS CONTACTYOURLOCALDISTRIBUTORORCALL1 800 834 4...

Reviews: