3.4 Vacuum measurement mode(VAC)
The high-voltage side channel is the vacuum test port of the
system.
!
When the pressure in the system is lower than the atmospheric
pressure (that is, the pressure display is less than 0), it will display the
vacuum degree of the system, otherwise it will display - - - -.
!
7.Exit the test result interface: long press the button.
Tip: The instrument has a built-in temperature compensation
function, which can perform the air tightness test of the
compression system. Leak analysis is performed by measuring
system pressure and ambient temperature over a period of time . If
the temperature compensation function is not required, it can be
turned off in the settings.
Note: This function needs to be purchased before it can be used, and
the purchase method is detailed in 4.4 of this article.
1. Connect the system to the high pressure side and open the high
pressure side valve.
2.Enter the vacuum mode: press the and button at the
same time , the upper left corner of the window will display VAC.
3. The product is connected to the vacuum pump and the system
under test.
4. Turn on the vacuum pump to extract the pressure in the system.
Figure 5 Test failure interface
Figure 6 Test result display interface
Summary of Contents for DMG-4B
Page 1: ...DMG 4B...
Page 2: ...1 1 1 1 2 DMG DMG 4B APP APP 2 3 4 5 5 4 1 7 8 9 6...
Page 3: ...1 3 1 4 1 2 3 4 5 6 1 2 3 1 5 1 4SAE 2 AAA 1 2 3 4 5 6 7 8 9...
Page 4: ...1 5 1 2 1 2 1 2 1 100psi 3 10 1 2...
Page 6: ...1 2 3 3 4 5 2 110 x 75 x 135mm 700g 10 50 14 122 20 60 4 140...
Page 7: ...1 2 1 100psi 3 4 1 1 3 3 1 3 2...
Page 8: ...4 4 1 3 APP 4 5 3 2 2 2 3 3...
Page 9: ...6 4 5 6 3 4 5...
Page 10: ...7 4 4 1 2 3 4 0 5 7 6 3 4...
Page 11: ...APP 4 4 1 APP 2 3 4 8 3000 5 6 7 7 3 5...
Page 12: ...9 8 9 3 6 SET 1 2 3 6 1...
Page 13: ...3 6 2 3 6 3 1 2 1 15 2 OFF 5 15 30 60min OFF 3 6 4 3 6 5 1 2 1 2 SN MAC 3 SN APP 1 2...
Page 15: ...4 4 4 5...
Page 16: ...4 6 18 APP 1 APP 2 3 1 APP 2 3 4 APP 1 APP...
Page 17: ...5 5 1 2 3 10 10...
Page 18: ...5 2 5 3 5 4 1 2 2 AAA 3 1 2 1 1 1 1 1 2 1 2 OL...
Page 19: ...English DMG 4B Digital Manifold Gauge User Manual...
Page 37: ...5 4 Packaging Accessories Product Host 1 Tool kit 1 Manual 1 Packing box 1...
Page 38: ...Deutsch DMG 4B 2 Wege Monteurhilfe Bedienungsanleitung...
Page 56: ...5 4 Verpackungszubeh r Produkthost 1 Werkzeugsatz 1 Manuell 1 Verpackungsbox 1...
Page 57: ...Fran ais Mode d emploi DMG 4B Manifold lectronique 2 voies...